Absolute Measurement Examples
6
Agilent E5505A User’s Guide
149
3
When you have completed these operations, click the
Close
button
.
Setup considerations for the free-running RF oscillator measurement
Measurement noise floor
The signal amplitude at the test set’s R input (Signal Input) port sets the
measurement noise floor level. Use
page 150 to determine the amplitude required to provide a noise floor level
that is below the expected noise floor of your DUT. (The Checking the Beatnote
procedure in this section will provide you with an opportunity to estimate the
measurement noise floor that your DUT will provide.)
Figure 100
Selecting loop suppression verification
e5505a_user_select_loop
24 Jun 04 rev 3
Summary of Contents for E1420B
Page 24: ...24 Agilent E5505A User s Guide ...
Page 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Page 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Page 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Page 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Page 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Page 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Page 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Page 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Page 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Page 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Page 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...