Preventive Maintenance
21
Agilent E5505A User’s Guide
455
•
Inspect the connectors before connection; look for dirt, nicks, and other
signs of damage or wear. A bad connector can ruin the good connector
instantly.
•
Clean dirty connectors. Dirt and foreign matter can cause poor electrical
connections and may damage the connector.
•
Minimize the number of times you bend cables.
•
Never bend a cable at a sharp angle.
•
Do not bend cables near the connectors.
•
If any of the cables will be flexed repeatedly, buy a back-up cable. This will
allow immediate replacement and will minimize system down time.
Before connecting the cables to any device:
•
Check all connectors for wear or dirt.
•
When making the connection, torque the connector to the proper value.
Proper Connector Torque
•
Provides more accurate measurements
•
Keeps moisture out of the connectors
•
Eliminates radio frequency interference (RFI) from affecting your
measurements
The torque required depends on the type of connector. Refer to
.
Do not overtighten the connector.
Never exceed the recommended torque when attaching cables.
Connector Wear and Damage
Look for metal particles from the connector threads and other signs of wear
(such as discoloration or roughness). Visible wear can affect measurement
accuracy and repeatability. Discard or repair any device with a damaged
connector. A bad connector can ruin a good connector on the first mating. A
magnifying glass or jeweler’s loupe is useful during inspection.
Table 75
Proper Connector Torque
Connector
Torque cm-kg
Torque N-cm
Torque in-lbs
Wrench P/N
Type-N
52
508
45
hand tighten
3.5 mm
9.2
90
8
8720-1765
SMA
5.7
56
5
8710-1582
Summary of Contents for E1420B
Page 24: ...24 Agilent E5505A User s Guide ...
Page 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Page 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Page 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Page 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Page 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Page 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Page 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Page 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Page 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Page 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Page 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...