Advanced Software Features
Ignore-Out-Of-Lock Mode
The Ignore Out Of Lock test mode enables all of the troubleshooting mode
functions, plus it causes the software to not check for an out-of-lock condition
before or during a measurement. This allows you to measure sources with high
close-in noise that normally would cause an out-of-lock condition and stop the
measurement. When Ignore Out Of Lock is selected, the user is responsible for
monitoring phase lock. This can be accomplished using an oscilloscope
connected to the test set Aux. Monitor port to verify the absence of a beatnote
and monitor the dc output level.
•
When Ignore Out Of Lock is selected, the test system does not verify the
phase lock of the measurement. The user must ensure that the
measurement maintains phase lock during the measurement.
Summary of Contents for E1420B
Page 24: ...24 Agilent E5505A User s Guide ...
Page 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Page 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Page 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Page 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Page 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Page 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Page 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Page 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Page 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Page 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Page 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...