AM Noise Measurement Fundamentals
11
Agilent E5505A User’s Guide
293
3
Measure the monitor output voltage on the AM detector with an
oscilloscope or voltmeter. Locate the diode detector’s DC voltage along the
bottom of the AM sensitivity graph (
). Moving up to the diagonal
calibration line and over, the equivalent phase detector constant can then
be read from the left side of the graph. The measured data will be plotted as
single-sideband AM noise in dBc/Hz.
4
Measure noise data and interpret the results.
Figure 226
Modulation sideband calibration setup
DUT
E5505a_mod_sideband_cal
02 Mar 04 rev 1
AM detector
Noise
input
Test set
DVM or
oscilloscope
Diode voltage
monitor output
NO T E
The quadrature meter should be at zero volts due to the blocking capacitor at the AM
detector’s output.
Summary of Contents for E1420B
Page 24: ...24 Agilent E5505A User s Guide ...
Page 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Page 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Page 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Page 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Page 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Page 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Page 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Page 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Page 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Page 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Page 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...