Agilent E5505A User’s Guide
10
FM Discriminator Measurement Examples
7
Choose the
Sources
tab from the
Define Measurement
window.
a
Enter the carrier (center) frequency of your DUT (5 MHz to 1.6 GHz).
Enter the same frequency for the detector input frequency. See
.
.
8
Choose the
Cal
tab from the
Define Measurement
window.
9
Select
Derive constant from FM rate and deviation
as the calibration
method.
10
Set the
FM Rate
to 20 kHz and
FM Deviation
to 10 kHz, which are the
recommended FM rate and deviation. See
Figure 202
Enter frequencies in Source tab
e5505a_user_enter_source_tab
25 Jun 04 rev 2