Residual Measurement Examples
8
Agilent E5505A User’s Guide
237
Making the measurement
Calibrate the measurement using measured ± DC peak voltage
Refer to
, “Residual Measurement Fundamentals for more
information about residual phase noise measurements calibration types.
Procedure
1
Using
and
on page 238 as guides, connect the circuit
and tighten all connections.
2
Measure the power level that will be applied to the Signal Input port of the
test set phase detector.
shows the acceptable amplitude ranges for
the test set phase detectors.
Table 37
Acceptable amplitude ranges for the phase detectors
Phase Detector
50 kHz to 1.6 GHz
1.2 to 26.5 GHz
1
1 Phase Noise Test Set Options 001 and 201
Ref Input (L Port)
Signal Input (R Port)
Ref Input (L Port)
Signal Input (R Port)
+
15 dBm
to
+
23 dBm
0 dBm
to
+
23 dBm
+
7 dBm
to
+
10 dBm
0 dBm
to
+
5 dBm
NO T E
Connecting an oscilloscope to the monitor port is recommended because the signal can
then be viewed to give visual confidence in the signal being measured. See
and
Summary of Contents for E1420B
Page 24: ...24 Agilent E5505A User s Guide ...
Page 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Page 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Page 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Page 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Page 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Page 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Page 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Page 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Page 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Page 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Page 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...