
382
Agilent E5505A User’s Guide
16
Reference Graphs and Tables
124: FM Dly equalizer
This special function allows you to turn off FM delay equalizer circuitry. When
[ON] (The preset condition), 30
µ
sec of group delay is added to the FM
modulated signal to get better FM frequency response.
You may want to turn [OFF] the FM Delay Equalizer circuitry when the signal
generator is used as the VCO in a phase-locked loop application to reduce
phase shift, of when you want to extend the FM bandwidth to
200 kHz. When [OFF], FM Indicator Accuracy is worse for rates of 1-5 kHz and
better beyond 30 kHz. Refer to the 8643A/8644B User’s Guide for specific
details.
Summary of Contents for E1420B
Page 24: ...24 Agilent E5505A User s Guide ...
Page 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Page 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Page 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Page 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Page 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Page 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Page 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Page 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Page 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Page 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Page 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...