212
Agilent E5505A User’s Guide
7
Residual Measurement Fundamentals
8
Once you have attained quadrature, you are ready to proceed with the
measurement.
Figure 151
Adjust for quadrature
e5505a_user_adjust_quad
24 Jun 04 rev 3
NO T E
For the system to accept the adjustment to quadrature, the meter must be
within
±
2 mV to
±
4 mV.
Figure 152
Measuring power at phase detector reference input port
Power
splitter
Phase
detector
Source
E5505a_pwr_phase_det_ref
27 Feb 04 rev 1
Optional line
stretcher
Power meter or
spectrum
analyzer
Test set
Signal
input
Ref input