FM Discriminator Fundamentals
9
Agilent E5505A User’s Guide
247
Optimum sensitivity
If measurements are made such that the offset frequency of interest (
) is
<1/2
the sin(x)/x term can be ignored and the transfer response can be
reduced to
where
is the discriminator constant.
The reduced transfer equation implies that a frequency discriminator’s system
sensitivity can be increased simply by increasing the delay
or by increasing
the phase detector constant
. This assumption is not completely correct.
is dependent on the signal level provided by the delay line and cannot exceed a
device dependent maximum. This maximum is achieved when the phase
detector is operating in
compression
1
. Increasing the delay
will reduce the
signal level out of the delay line often reducing the sensitivity of the phase
detector. Optimum system sensitivity is obtained in a trade-off between delay
and attenuation.
Sensitivity = K
L
V
in
LX(10)
−
LZ/20
Where K
L
is the phase detector efficiency, V
in
is the signal voltage into the
delay line, LX (dB) is the sensitivity provided by the delay line and LZ is the
attenuation of the delay line. Taking the derivative with respect to the length L
to find the maximum of this equation results in
LZ = 8.7 dB of attenuation
The optimum sensitivity of a system with the phase detector operating out of
results from using a length of coaxial line that has 8.7 dB of attenuation.
One way to increase the sensitivity of the discriminator when the phase
detector is out of compression is to increase the signal into the delay line. This
can be accomplished with an RF amplifier before the signal splitter. The noise
of the RF amplifier will not degrade the measurement if the two-port noise of
the amplifier is much less than the noise of the DUT. However, some
attenuation may be needed in the signal path to the reference input to the
double-balanced mixer (phase detector) to protect it from excessive power
levels.
If the amplifier signal puts the phase detector into compression,
is at its
maximum and system sensitivity is now dependent on the length of the delay
. For maximum sensitivity more delay can be added until the signal level out
of the delay line is 8.7 dB below the phase detector compression point.
The following example illustrates how to choose a delay line that provided the
optimum sensitivity given certain system parameters. (See
1
Compression
: The level of the output signal at which the gain of a device is reduced by a specific amount,
usually expressed in decibels (dB), as in the 1 dB compression point.
fm
πτ
d
∆
V fm
( )
Kd
∆
f fm
( )
K
φπτ
d
∆
f fm
( )
=
=
Kd
τ
d
K
φ
K
φ
τ
d
K
φ
τ
d
Summary of Contents for E1420B
Page 24: ...24 Agilent E5505A User s Guide ...
Page 46: ...46 Agilent E5505A User s Guide 2 Introduction and Measurement ...
Page 128: ...128 Agilent E5505A User s Guide 5 Absolute Measurement Fundamentals ...
Page 226: ...226 Agilent E5505A User s Guide 7 Residual Measurement Fundamentals ...
Page 322: ...322 Agilent E5505A User s Guide 13 Baseband Noise Measurement Examples ...
Page 340: ...340 Agilent E5505A User s Guide 14 Evaluating Your Measurement Results ...
Page 358: ...358 Agilent E5505A User s Guide 15 Advanced Software Features ...
Page 386: ...386 Agilent E5505A User s Guide 16 Reference Graphs and Tables ...
Page 392: ...392 Agilent E5505A User s Guide 17 System Specifications ...
Page 404: ...404 Agilent E5505A User s Guide 18 System Interconnections ...
Page 452: ...452 Agilent E5505A User s Guide 20 PC Digitizer Performance Verification ...
Page 466: ...466 Agilent E5505A User s Guide 21 Preventive Maintenance ...