![Agilent Technologies B1500A User Manual Download Page 89](http://html.mh-extra.com/html/agilent-technologies/b1500a/b1500a_user-manual_2867697089.webp)
Agilent B1500A User’s Guide, Edition 10
3- 15
Installation
Installing Accessories
Table 3-3
To Connect Connector Plate
Cables
Connector
Plate
B1500A Terminals
Agilent 16493J Interlock Cable, 3 m or 1.5 m
Intlk
Interlock
Agilent 16493L GNDU Cable, 3 m or 1.5 m
GNDU
GNDU
Agilent 16493K
a
Kelvin Triaxial Cable, 3 m or
1.5 m, for the Kelvin connection
16495F provides the inputs 1 to 12 (6 Kelvin).
16495H provides the inputs 1 to 6 (3 Kelvin).
16495J provides the inputs 1 to 8 (4 Kelvin).
a. Use Agilent 16494B Kelvin Triaxial Cable to connect to Agilent E5250A
Low Leakage Switch Mainframe or the connector plate for Agilent 4142B
Modular DC Source/Monitor by the Kelvin connection.
odd number
SMU
Force
even number
Sense
Agilent 16494A Triaxial Cable, 3 m or 1.5 m,
for non-Kelvin connection
16495H provides the inputs 1 to 6.
16495J provides the inputs 1 to 8.
odd number
or
even number
SMU
Force
Agilent 16493P SPGU Cable, 3 m or 1.5 m
16495H provides the inputs 7 to 12.
16495J provides the inputs 9 to 12.
odd number
or
even number
SPGU
Output 1
Output 2
Agilent N1300A CMU Cable
b
, 3 m or 1.5 m
16495H provides the inputs 7 to 12.
16495J provides the inputs 9 to 12.
b. To extend the four-terminal pair (4TP) connection to the connector plate
inputs, connect the Hcur and Hpot cables together and the Lcur and Lpot
cables together by using the BNC T-type adapter (female-female-male, Agi-
lent part number 1250-2405), then connect them to the connector plate inputs.
To extend the 4TP connection to the DUT interface, connect the cables to the
connector plates input connectors individually, and connect the Hcur and
Hpot terminals together and the Lcur and Lpot terminals together on the DUT
interface. See Figure 3-17.
It is ok to leave open the ground wire extended from the CMU cable.
odd number
or
even number
CMU
Hcur/Hpot
Lcur/Lpot
Summary of Contents for B1500A
Page 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Page 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Page 23: ...1 Getting Started ...
Page 41: ...2 Introduction ...
Page 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Page 75: ...3 Installation ...
Page 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Page 133: ...4 Specifications ...
Page 179: ...5 Accessories and Options ...