2- 16
Agilent B1500A User’s Guide, Edition 10
Introduction
Rear View
WARNING
There are potentially hazardous voltages (200 V for HPSMU, and 100 V for
MPSMU/HRSMU/ASU) present at the Force, Sense, and Guard terminals of
Agilent B1500A. To prevent electrical shock, the following safety precautions
must be observed during the use of the B1500A.
a. Use a three-conductor AC power cable to connect the cabinet (if used)
and the B1500A to an electrical ground (safety ground).
b. If you do
not
use the 16442B test fixture (B1500A-A5F), you must install
and connect an interlock circuit that opens the Interlock terminal when
the shielding box access door is opened.
c. Confirm periodically that the interlock function works normally.
d. Before touching the connections on the Force, Guard, and Sense
terminals, turn the B1500A off and discharge any capacitors. If you do
not
turn the B1500A off, complete
all
of the following items, regardless
of the B1500A settings.
•
Press the
Stop
key to set the module output off.
•
Confirm that the High Voltage indicator is not lit.
•
Open the shielding box access door (open the Interlock terminal).
•
Discharge any capacitors connected to an SMU.
e. Warn persons working around the B1500A about dangerous conditions.
E. CMU module
The multi frequency capacitance measurement unit (MFCMU) has four coaxial
connectors, Lcur, Lpot, Hpot, and Hcur, for the four-terminal pair connection.
The Hcur and Hpot have to be connected together at a terminal of the device
under test (DUT). Also the Lcur and Lpot have to be connected together at the
other terminal of the DUT. Then, the guard lines of four-terminal pair must be
connected together.
CAUTION
Do not apply voltage more than ±25 V to the MFCMU input terminals. Failure to
heed this caution may result in damage to the module.
Summary of Contents for B1500A
Page 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Page 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Page 23: ...1 Getting Started ...
Page 41: ...2 Introduction ...
Page 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Page 75: ...3 Installation ...
Page 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Page 133: ...4 Specifications ...
Page 179: ...5 Accessories and Options ...