Pulse range and pulse parameter
Item
Value
a
a. Unless otherwise stated, all specifications assume a 50
termination.
Frequency range
0.1 Hz to 33 MHz
Pulse period
Programmable range
30 ns to 10 s
Resolution
10 ns
Minimum
100 ns (Vamp
10 V)
Accuracy
± 1 % (± 0.01 %
b
)
b. Typical minimum. This is the supplemental information.
Pulse width
Programmable range
10 ns to (period
-
10 ns)
Resolution
2.5 ns (Tr and Tf
8
s)
10 ns (Tr or Tf > 8
s)
Minimum
50 ns (25 ns, typical) at Vamp
10 V
Accuracy
± (3 % + 2 ns)
Transition time (leading time Tr
and trailing time Tf)
c
c. The time from 10 % to 90 % of Vamp which is the amplitude of output pulse.
Programmable range
8 ns to 400 ms (see below for the minimum
transition time)
Resolution
2 ns (Tr and Tf
8
s)
8 ns (Tr or Tf > 8
s)
Minimum (typical)
< 15 ns (Vamp
10 V)
Minimum
20 ns (Vamp
10 V)
30 ns (Vamp
20 V)
60 ns (Vamp > 20 V)
Accuracy
-
5 % to 5 % + 10 ns (Vamp
10 V)
-
5 % to 5 % + 20 ns (Vamp
20 V)
Output relay switching time
d
d. The time it takes the open state relay to open or close.
Open/close
< 100
s
Summary of Contents for B1500A
Page 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Page 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Page 23: ...1 Getting Started ...
Page 41: ...2 Introduction ...
Page 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Page 75: ...3 Installation ...
Page 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Page 133: ...4 Specifications ...
Page 179: ...5 Accessories and Options ...