![Agilent Technologies B1500A User Manual Download Page 149](http://html.mh-extra.com/html/agilent-technologies/b1500a/b1500a_user-manual_2867697149.webp)
Agilent B1500A User’s Guide, Edition 10
4- 17
Specifications
Hardware
MFCMU
MFCMU: Multi Frequency Capacitance Measurement Unit
•
Measurement functions
•
Measurement parameters: Cp-G, Cp-D, Cp-Q, Cp-Rp, Cs-Rs, Cs-D, Cs-Q,
Lp-G, Lp-D, Lp-Q, Lp-Rp, Ls-Rs, Ls-D, Ls-Q, R-X, G-B, Z-
, Y-
•
Ranging: Auto or fixed
•
Measurement terminal: Four-terminal pair configuration, four BNC (female)
connectors
•
Cable length: 1.5 m or 3 m, automatic identification of accessories
•
Test signal
•
Frequency
Range: 1 kHz to 5 MHz
Resolution: 1 mHz (minimum)
Accuracy: ± 0.008 %
•
Output signal level
Range: 10 mVrms to 250 mVrms
Resolution: 1 mVrms
Accuracy:
± (10.0 % + 1 mVrms) at the measurement port of the MFCMU
± (15.0 % + 1 mVrms) at the measurement port of the MFCMU cable (1.5 m
or 3.0 m)
•
Output impedance: 50
, typical
•
Signal level monitor
Range: 10 mVrms to 250 mVrms
Accuracy (open load):
± (10.0 % of r 1 mVrms) at the measurement port of the MFCMU
± (15.0 % of r 1 mVrms) at the measurement port of the MFCMU
cable (1.5 m or 3.0 m)
Summary of Contents for B1500A
Page 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Page 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Page 23: ...1 Getting Started ...
Page 41: ...2 Introduction ...
Page 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Page 75: ...3 Installation ...
Page 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Page 133: ...4 Specifications ...
Page 179: ...5 Accessories and Options ...