![Agilent Technologies B1500A User Manual Download Page 104](http://html.mh-extra.com/html/agilent-technologies/b1500a/b1500a_user-manual_2867697104.webp)
3- 30
Agilent B1500A User’s Guide, Edition 10
Installation
Mounting Connectors
Installing the interlock circuit
Install the interlock circuit as follows:
1. Mount two mechanical switches on your shielding box, so that the switches
close when the door of the shielding box is closed, and open when the door is
opened. For the dimensions of the switch, see Figure 3-10 and Figure 3-11
below.
2. Mount an LED on your shielding box. For the dimensions of the LED, see
3. Use wire to connect the two switches in series between pin number 1 and 2 (or
3) of the interlock connector. See Figure 3-8.
4. Use wire to connect the LED between pin number 4 and 5 (or 6) of the interlock
connector. See Figure 3-8.
If Agilent B1500A Interlock connector is connected to the interlock circuit, Agilent
B1500A SMU
cannot
force more than ± 42 V when the door is open. When the door
is closed, it can force more than ± 42 V.
When more than ± 42 V is forced from an SMU, the LED lights to indicate
high
voltage output
.
Figure 3-9
Dimensions of the LED (Agilent part number 1450-0641)
8QLWV PP
$QRGH
&DWKRGH
Summary of Contents for B1500A
Page 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Page 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Page 23: ...1 Getting Started ...
Page 41: ...2 Introduction ...
Page 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Page 75: ...3 Installation ...
Page 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Page 133: ...4 Specifications ...
Page 179: ...5 Accessories and Options ...