![Agilent Technologies B1500A User Manual Download Page 157](http://html.mh-extra.com/html/agilent-technologies/b1500a/b1500a_user-manual_2867697157.webp)
Agilent B1500A User’s Guide, Edition 10
4- 25
Specifications
Hardware
HV-SPGU
HV-SPGU: High Voltage Semiconductor Pulse Generator Unit
•
Number of output channels: 2 channels per module
•
Modes: pulse, constant, and free run
•
Standard pulse mode
•
Two level pulse
•
Three level pulse per one channel
•
Pulse period: 30 ns to 10 s
•
Delay range: 0 s to 9.99 s
•
Delay resolution: minimum 2.5 ns
•
Output count: 1 to 1,000,000
•
Voltage monitor minimum sampling period: 5
s
•
Trigger output
Level: TTL
Timing: Synchronized with pulse period
Trigger width: Pulse period
1/2 (pulse period
10
s), maximum 5
s (pulse
period > 10
s)
•
Pulse range and pulse parameter: See Table 4-15.
•
Pulse/DC output voltage and accuracy: See Table 4-16.
Figure 4-1
SPGU pulse setup parameters
Tperiod
Twidth
Tdelay
Vamp
Vpeak
Vbase
Tr
Tf
90%
90%
10%
10%