![Agilent Technologies B1500A User Manual Download Page 150](http://html.mh-extra.com/html/agilent-technologies/b1500a/b1500a_user-manual_2867697150.webp)
4- 18
Agilent B1500A User’s Guide, Edition 10
Specifications
Hardware
•
DC bias function
•
DC bias
Range: 0 to ± 25 V
Resolution: 1 mV
Accuracy: ± (0.5 % + 5.0 mV) at the measurement port of the MFCMU or
the MFCMU cable (1.5 m or 3.0 m)
•
Maximum DC bias current, supplemental information:
•
Output impedance: 50
, typical
•
DC bias monitor
Range: 0 to ± 25 V
Accuracy (open load): ± (0.2 % of r 10 mV) at the measurement
port of the MFCMU or the MFCMU cable (1.5 m or 3.0 m)
•
Sweep characteristics
•
Available sweep parameters: DC bias voltage, oscillator level, frequency
•
Sweep type: linear, log
•
Sweep mode: single, double
•
Sweep direction: up, down
•
Number of measurement points: maximum 1001
Impedance range (
)
Maximum DC bias current (A)
50
10
10
-3
100
10
10
-3
300
10
10
-3
1
10
3
1
10
-3
3
10
3
1
10
-3
10
10
3
100
10
-6
30
10
3
100
10
-6
100
10
3
10
10
-6
300
10
3
10
10
-6
Summary of Contents for B1500A
Page 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Page 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Page 23: ...1 Getting Started ...
Page 41: ...2 Introduction ...
Page 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Page 75: ...3 Installation ...
Page 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Page 133: ...4 Specifications ...
Page 179: ...5 Accessories and Options ...