![Agilent Technologies B1500A User Manual Download Page 66](http://html.mh-extra.com/html/agilent-technologies/b1500a/b1500a_user-manual_2867697066.webp)
2- 26
Agilent B1500A User’s Guide, Edition 10
Introduction
Measurement Units
Table 2-9
HRSMU Output Voltage Ranges and Resolutions
Table 2-10
HRSMU Output Current Ranges and Resolutions
Range
Output Value
Setting
Resolution
Maximum
Current
0.5 V
0
|V|
0.5 V
25
V
± 100 mA
2 V
0
|V|
2 V
100
V
± 100 mA
5 V
0
|V|
5 V
250
V
± 100 mA
20 V
0
|V|
20 V
1 mV
± 100 mA
40 V
0
|V|
20 V
2 mV
± 100 mA
20 V < |V|
40 V
2 mV
± 50 mA
100 V
0
|V|
20 V
5 mV
± 100 mA
20 V < |V|
40 V
5 mV
± 50 mA
40 V < |V|
100 V
5 mV
± 20 mA
Range
Output Value
Setting
Resolution
Maximum
Voltage
1 pA
a
a. Available when the E5288A Atto Sense and Switch Unit (ASU) is used.
0
|I|
1.15 pA
1 fA
± 100 V
10 pA
0
|I|
11.5 pA
5 fA
± 100 V
100 pA
0
|I|
115 pA
5 fA
± 100 V
1 nA
0
|I|
1.15 nA
50 fA
± 100 V
10 nA
0
|I|
11.5 nA
500 fA
± 100 V
100 nA
0
|I|
115 nA
5 pA
± 100 V
1
A
0
|I|
1.15
A
50 pA
± 100 V
10
A
0
|I|
11.5
A
500 pA
± 100 V
100
A
0
|I|
115
A
5 nA
± 100 V
1 mA
0
|I|
1.15 mA
50 nA
± 100 V
10 mA
0
|I|
11.5 mA
500 nA
± 100 V
100 mA
0
|I|
20 mA
5
A
± 100 V
20 mA < |I|
50 mA
5
A
± 40 V
50 mA < |I|
100 mA
5
A
± 20 V
Summary of Contents for B1500A
Page 1: ...Agilent Technologies Agilent B1500A Semiconductor Device Analyzer User s Guide ...
Page 22: ...Agilent B1500A User s Guide Edition 10 Contents ...
Page 23: ...1 Getting Started ...
Page 41: ...2 Introduction ...
Page 74: ...2 34 Agilent B1500A User s Guide Edition 10 Introduction Measurement Units ...
Page 75: ...3 Installation ...
Page 132: ...3 58 Agilent B1500A User s Guide Edition 10 Installation About Plug in Modules ...
Page 133: ...4 Specifications ...
Page 179: ...5 Accessories and Options ...