Chapter 2
2-79
System Verification and Performance Tests
Agilent 8753ES System Verification and Performance Tests
Procedure
Test Port Output Worst-Case 2nd Harmonic
1. Press
to set the test port power to +10 dBm. (Set the power
to +8 dBm for analyzers with Option 014 or 075.)
2. Press
. To set the frequency range:
If you do not have Option 006, press
.
If you have Option 006, press
.
3. Press
to set the IF bandwidth to 10 Hz.
4. Connect the equipment as shown in
.
Figure 2-44
Test Port Output Harmonics Test Setup
5. Press
.
6. After one sweep, press
to normalize the
trace.
7. Press
.
8. After one sweep, press
to get a better viewing of the trace.
9. Press .
10.Write the marker 1 value (which appears on the analyzer display) on the performance
test record. This is the worst-case test port output 2nd harmonic.
Preset
Power
10
x1
Start
16
M/
µ
Stop
1.5
G/n
Stop
3
G/n
Avg
IF BW
10
x1
Meas
Trans:REV S12 (A/R) INPUT PORTS A
Display
DATA
→
MEMORY DATA/MEM
System
HARMONIC MEAS HARMONIC SECOND
Scale Ref
AUTO SCALE
Marker Search
SEARCH:MAX
Summary of Contents for 8753ES
Page 14: ...Contents xiv Contents ...
Page 15: ...1 1 1 Service Equipment and Analyzer Options ...
Page 26: ...1 12 Chapter1 Service Equipment and Analyzer Options Service and Support Options ...
Page 27: ...2 1 2 System Verification and Performance Tests ...
Page 203: ...3 1 3 Adjustments and Correction Constants ...
Page 262: ...3 60 Chapter3 Adjustments and Correction Constants Sequences for Mechanical Adjustments ...
Page 263: ...4 1 4 Start Troubleshooting Here ...
Page 297: ...5 1 5 Power Supply Troubleshooting ...
Page 317: ......
Page 318: ...6 1 6 Digital Control Troubleshooting ...
Page 337: ...6 20 Chapter6 Digital Control Troubleshooting GPIB Failures ...
Page 338: ...7 1 7 Source Troubleshooting ...
Page 369: ...7 32 Chapter7 Source Troubleshooting Source Group Troubleshooting Appendix ...
Page 370: ...8 1 8 Receiver Troubleshooting ...
Page 381: ...8 12 Chapter8 Receiver Troubleshooting Troubleshooting When One or More Inputs Look Good ...
Page 382: ...9 1 9 Accessories Troubleshooting ...
Page 389: ...9 8 Chapter9 Accessories Troubleshooting Inspect the Error Terms ...
Page 390: ...10 1 10 Service Key Menus and Error Messages ...
Page 439: ...10 50 Chapter10 Service Key Menus and Error Messages Error Messages ...
Page 440: ...11 1 11 Error Terms ...
Page 451: ...11 12 Chapter11 Error Terms Error Correction ...
Page 452: ...12 1 12 Theory of Operation ...
Page 482: ...13 1 13 Replaceable Parts ...
Page 487: ...13 6 Chapter13 Replaceable Parts Ordering Information Figure 13 1 Module Exchange Procedure ...
Page 500: ...Chapter 13 13 19 Replaceable Parts Replaceable Part Listings Figure 13 7 8753ET Cables Top ...
Page 502: ...Chapter 13 13 21 Replaceable Parts Replaceable Part Listings Figure 13 8 8753ES Cables Top ...
Page 512: ...Chapter 13 13 31 Replaceable Parts Replaceable Part Listings Figure 13 13 8753ES Cables Front ...
Page 544: ...14 1 14 Assembly Replacement and Post Repair Procedures ...
Page 550: ...Chapter 14 14 7 Assembly Replacement and Post Repair Procedures Covers Figure 14 2 Covers ...
Page 597: ...14 54 Chapter14 Assembly Replacement and Post Repair Procedures Post Repair Procedures ...