Chapter 1
19
General Information
Performing the Operator’s Check
NOTE
This is an 80% confidence test only. A test set could pass this simple test
and yet still not function properly. For more complete testing, see
“Performance Tests” on page 45
.
If a more complete test is desired, repeat the above procedure, but this
time terminate each port being tested with a known good 75
Ω
load
(with a return loss of better than 40 dB). The resulting return loss
should be a maximum of 15 dB.