Chapter 4
53
Service
Troubleshooting
Troubleshooting the Front Panel Board
Turn the instrument power on and check the following:
1. Check the condition of each of the switching paths by issuing commands to switch each
of the paths to either the reflection or the transmission path. Ensure that the LED
indicates the appropriate path.
2. If the LED indicates a wrong path, the problem can lie with either the front panel
board or the main switch board. Determine where the problem is by measuring the RF
path.
3. If the LED does not display the proper path, check to see if the RF path has indeed
been switched. If the problem lies with the front panel board, replace it.
4. Ensure that the front panel washers between the board and the front panel assembly
are present. Missing washers can cause erratic LED behavior.
Troubleshooting the Controller and Switch Driver Boards
Turn the instrument power on. Check the condition of each of the switching paths by
issuing commands to switch each of the paths to either the reflection or the transmission
path. Check each of the RF paths for connection. If an RF path is not connected to the
necessary port or terminated in 75
Ω
, replace the controller and switch driver board.
Refer to
Figure 4-1 on page 54
for a block diagram of the major components and the
switching paths of the 87075C Option H08.
Summary of Contents for 87075C Option H08
Page 1: ...Agilent Technologies 87075C Option H08 User s and Service Guide ...
Page 2: ......
Page 6: ...4 User s and Service Guide Supplement ...
Page 9: ...7 1 General Information ...
Page 22: ...20 Chapter1 General Information Performing the Operator s Check ...
Page 23: ...21 2 Controlling the Test Set and Making Measurements ...
Page 38: ...36 Chapter2 Controlling the Test Set and Making Measurements Calibrating the Test System ...
Page 39: ...37 3 Front Rear Panel ...
Page 45: ...43 4 Service ...
Page 60: ...58 Chapter4 Service Theory of Operation ...