![Agilent Technologies 87075C Option H08 User'S And Service Manual Download Page 13](http://html.mh-extra.com/html/agilent-technologies/87075c-option-h08/87075c-option-h08_users-and-service-manual_2867954013.webp)
Chapter 1
11
General Information
Requirements
Requirements
Electrical
The alternating-current (AC) power that is supplied to the test set
must meet the following requirements:
Voltage: 90 to 250 Vac
Frequency: 48 to 66 Hz
Available power: 40 watts minimum
If the available AC line voltage is outside the 90 to 250 Vac range, an
autotransformer that provides third wire continuity to earth ground
may be used.
Environmental
Operating Environment
Indoor use only
Operating temperature: 0 to 55 °C
Maximum relative humidity: 80 percent for temperatures up to 31 °C
decreasing linearly to 50 percent relative humidity at 40 °C
Altitude: up to 15,000 feet (4,572 meters)
Enclosure protection: IP 20, according to IEC 529
CAUTION
This product is designed for use in INSTALLATION CATEGORY II,
and POLLUTION DEGREE 2, per IEC 101 and 664 respectively.
Non-Operating Storage Conditions
Temperature: –40 °C to +70 °C
Humidity: 0 to 90 percent relative at +65 °C (non-condensing)
Altitude: 0 to 15,240 meters (50,000 feet)
Weight
Net: Approximately 7 kg
Shipping: Approximately 20 kg
Summary of Contents for 87075C Option H08
Page 1: ...Agilent Technologies 87075C Option H08 User s and Service Guide ...
Page 2: ......
Page 6: ...4 User s and Service Guide Supplement ...
Page 9: ...7 1 General Information ...
Page 22: ...20 Chapter1 General Information Performing the Operator s Check ...
Page 23: ...21 2 Controlling the Test Set and Making Measurements ...
Page 38: ...36 Chapter2 Controlling the Test Set and Making Measurements Calibrating the Test System ...
Page 39: ...37 3 Front Rear Panel ...
Page 45: ...43 4 Service ...
Page 60: ...58 Chapter4 Service Theory of Operation ...