28
Chapter 2
Controlling the Test Set and Making Measurements
Calibrating the Test System
Figure 2-3
Calibrating the Test System
For the best accuracy perform a full two-port calibration between the
test set PORT 1 and PORT 3, and again between PORT 1 and PORT 5.
Save the calibrations as instrument states. Refer to the network
analyzer user’s guide for information on calibrations and saving
instrument states.
CAUTION
When performing a full two-port calibration and making subsequent
measurements, you must use the transfer switch internal to the
8753D/E analyzer to change the RF signal path direction. Do not use
the test set to change the RF signal path direction when you are
entering a full two-port calibration. Doing so will render the calibration
invalid.
NOTE
The SelfCal feature (described in the 87050E /87075C User’s and
Service Guide) is not available when using the Option H08 with the
8753D/E analyzer.
Making Measurements
The following examples assume that you are using a parallel port
connection with an 8753D/E, with the test set’s parallel address set to
“0”. See
“Setting the Test Set Address Switch” on page 17
for
information on setting the test set’s address.
Summary of Contents for 87075C Option H08
Page 1: ...Agilent Technologies 87075C Option H08 User s and Service Guide ...
Page 2: ......
Page 6: ...4 User s and Service Guide Supplement ...
Page 9: ...7 1 General Information ...
Page 22: ...20 Chapter1 General Information Performing the Operator s Check ...
Page 23: ...21 2 Controlling the Test Set and Making Measurements ...
Page 38: ...36 Chapter2 Controlling the Test Set and Making Measurements Calibrating the Test System ...
Page 39: ...37 3 Front Rear Panel ...
Page 45: ...43 4 Service ...
Page 60: ...58 Chapter4 Service Theory of Operation ...