Chapter 4
47
Service
Performance Tests
Isolation
NOTE
Isolation needs to only be measured on adjacent ports. Two 75
Ω
loads
are required for this test.
1. Perform a full two-port calibration from 3 MHz to 1.3 GHz at the
ends of two cables attached to the two test ports of the network
analyzer. Make sure the calibration is active.
2. Connect a 75
Ω
load to both the Reflection and Transmission ports of
the 87075C Option H08.
3. Connect the two cables that are attached to the network analyzer, to
ports 1 and 2 of the 87075C Option H08. The exact order doesn’t
matter.
4. Using the “CONTROL” program provided, select reflection Port 1
and transmission Port 2, e.g. “1,2” (you could also use “2,1”). Set the
8753D/E to measure the transmission. Verify > 60 dB of isolation.
5. Repeat steps 3 and 4 for the next two adjacent ports (2 and 3).
Repeat again for ports 3 and 4, then 4 and 5, and so on, ending with
ports 7 and 8.
Summary of Contents for 87075C Option H08
Page 1: ...Agilent Technologies 87075C Option H08 User s and Service Guide ...
Page 2: ......
Page 6: ...4 User s and Service Guide Supplement ...
Page 9: ...7 1 General Information ...
Page 22: ...20 Chapter1 General Information Performing the Operator s Check ...
Page 23: ...21 2 Controlling the Test Set and Making Measurements ...
Page 38: ...36 Chapter2 Controlling the Test Set and Making Measurements Calibrating the Test System ...
Page 39: ...37 3 Front Rear Panel ...
Page 45: ...43 4 Service ...
Page 60: ...58 Chapter4 Service Theory of Operation ...