4349B
The
typical
pull-up
resistor
values
are:
Pull-up
V
oltage
Pull-up
Resistor
Agilent
P
art
Number
5
V
0757-0278
(1.78
k)
9
V
0757-0279
(3.16
k)
12
V
0698-3154
(4.22
k)
15
V
0757-0438
(5.11
k)
EXT
DCV2
EXT
TRIG
KEY
LOCK
SW1-1
SW1-2
SW1-3
5
to
6
V
Close
Close
Close
6
to
9
V
Close
Open
Close
9
to
15
V
Open
Close
Open
8.
Reinstall
the
A1
main
board,
the
A2
board,
and
the
cover
.
B-10
Handler
Interface
Installation