4349B
:FETCh?
FETCh?
Query
:FETCh?
The
FETCh?
query
retrieves
the
measurements
taken
by
the
INITiate
command
and
places
them
into
the
4349B's
output
buer
.
(Query
only)
Query
response
is
,
<stat1>,<data1>,<comp1> ,<stat2>,< data2>,<comp2> ,<stat3>,<data3> ,<comp3>,
<stat4>,<data4>,<comp4>
Where
,
<stat1-4>
measurement
status
of
each
channel
0
:
normal
1
:
overload
2
:
no-contact
<data1-4>
measurement
data
of
each
channel
<comp1-4>
comparison
result
of
each
channel
(no
output
when
the
comparator
function
is
OFF)
1
:
In
2
:
High
4
:
Low
8
:
No-Contact
F
ormat
is
specied
by
FORMat
subsystem.
F
or
more
information,
refer
\T
o
Trigger
a
Measurement"
in
Chapter
4.
GPIB
Reference
5-17