4349B
3.
Press
the
3458A
Multimeter's
4
DCV
5
to
set
the
measurement
mode
to
DC
voltage
.
4.
Press
three
times
to
set
the
measurement
time
to
10
ms
.
The
measurement
time
setting
is
indicated
by
the
Meas
Time
annunciator(
9
).
5.
Press
to
set
the
trigger
mode
to
Manual.
Trigger
mode
is
indicated
by
the
Trigger
annunciator(
9
).
6.
Set
the
measurement
range
to
100
A
using
the
following
procedure
.
a.
Press
to
display
the
measurement
range
setup
menu.
b.
Select
100
A
using
or
and
press
.
7.
Record
the
multimeter
reading
into
the
performance
test
record.
8.
Press
to
select
channel
2.
9.
Connect
the
multimeter
input
to
the
4349B 's
channel
2
input
instead
of
the
channel
1
input.
10.
T
est
channel
2
using
the
steps
6
to
7.
11.
T
est
channel
3
and
channel
4
using
the
same
procedure
.
If
the
4349B
is
equipped
with
Option
001
(Two
T
est
Channels),
skip
this
step
.
Ammeter
Input
Resistance
T
est.
12.
Record
the
16340A
RC
Box's
10
4
calibration
value
and
the
ammeter
oset
voltage
test
result
on
the
calculation
sheet.
13.
Set
up
the
equipment
as
shown
in
Figure
9-2.
If
the
DC
V
oltage
Calibrator
listed
in
T
able
9-2
is
available
,
setup
the
equipment
as
shown
in
Figure
9-3.
Figure
9-2.
Ammeter
Input
Resistance
T
est
Setup
Using
DC
P
ower
Supply
9-4
Maintenance
Summary of Contents for 4349B
Page 1: ......
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 20: ......
Page 42: ...4349B Figure 1 1 Power Cable Supplied 1 10 Getting Started ...
Page 66: ......
Page 88: ......
Page 104: ......
Page 146: ......
Page 160: ......
Page 176: ...4349B Figure 8 9 10 Additional Error Measurement Time 400 msec 8 14 General Information ...
Page 212: ...4349B Figure B 3 Handler Interface Input Signal Diagram B 4 Handler Interface Installation ...
Page 216: ...4349B Figure B 4 A1 Main Board B 8 Handler Interface Installation ...
Page 228: ......