4349B
Figure
9-3.
Ammeter
Input
Resistance
T
est
Setup
Using
DC
V
oltage
Calibrator
14.
Set
the
RC
Box
resistor
to
10
4
.
15.
Set
the
DC
power
supply
(voltage
calibrator)
to
1
V
.
16.
Record
the
multimeter
reading
on
the
calculation
sheet
on
the
L
V
oltage
line
.
17.
Disconnect
the
Dual
Banana-BNC(m)
cable
from
A
of
Figure
9-2,
and
connect
it
to
B
.
If
the
voltage
calibrator
is
used,
skip
this
step
.
18.
Record
the
multimeter
reading
(voltage
calibrator
setting)
on
the
calculation
sheet
on
the
H
V
oltage
line
.
19.
Calculate
the
test
result
according
to
the
calculation
sheet,
and
record
the
result
into
the
performance
test
record.
20.
Remove
the
triaxial
cable
from
the
4349B's
channel
1
input
and
connect
it
to
channel
2
input.
21.
T
est
channel
2
using
the
steps
13
to
19.
22.
T
est
channel
3
and
channel
4
using
the
same
procedure
.
If
the
4349B
is
equipped
with
Option
001
(Two
T
est
Channels),
skip
this
step
.
Maintenance
9-5
Summary of Contents for 4349B
Page 1: ......
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 20: ......
Page 42: ...4349B Figure 1 1 Power Cable Supplied 1 10 Getting Started ...
Page 66: ......
Page 88: ......
Page 104: ......
Page 146: ......
Page 160: ......
Page 176: ...4349B Figure 8 9 10 Additional Error Measurement Time 400 msec 8 14 General Information ...
Page 212: ...4349B Figure B 3 Handler Interface Input Signal Diagram B 4 Handler Interface Installation ...
Page 216: ...4349B Figure B 4 A1 Main Board B 8 Handler Interface Installation ...
Page 228: ......