Chapter 3: Testing Logic Analyzer Performance
Configure the Logic Analysis System
f
Ensure that the sampling speed is set to
250 MHz
in the Sampling
Options box.
NOTE:
If option 500 is not installed on the 16910/11A module, then 250 MHz will
be the only speed available.
g
Ensure that the Clock Mode is set to
Master
.
h
Set the clock mode to
Both Edges
.