42
Chapter 3: Testing Logic Analyzer Performance
Configure the Logic Analysis System
mainframe you can touch the touchscreen and drag across with your
finger.
f
Click (or touch) to select channels 2, 6, 10 and 14 as shown.
g
Drag the scroll bar all the way to the left and ensure that the activity
indicator shows activity on clock 1.
4
Set the sampling mode.
a
Select the
Sampling
tab of the Analyzer Setup window.
b
Select
State Mode
.
c
Set the Trigger Position to
100% Poststore
.
d
Set the Acquisition Depth to
128K
.
e
Clear the Timing Zoom check box to turn Timing Zoom off.