Pod Threshold Field
The pod threshold field is used to set a voltage level which the data must
cross before the analyzer recognizes and displays it as a change in logic
levels. You specify a threshold level for each pod. The level specified for the
master pod, which includes the clock, is also assigned to the clock.
TTL
When TTL is selected as the threshold level, the data signals must
cross +1.5 volts.
ECL
When ECL is selected as the threshold level, the data signals must
cross
−
1.3 volts.
USER
When USER is selected as the threshold level, the data signals
must cross a user selectable value between
−
5.0 volts to +5.0 volts.
Pod Threshold Field
Pod threshold field
Pod threshold selection
menu
The Format Menu
Pod Threshold Field
3–8
Summary of Contents for 16517A
Page 3: ...The Agilent Technologies 16517A 4 GSa S Timing and 1 GSa S State Logic Analyzer ii ...
Page 5: ...iv ...
Page 11: ...Contents 6 ...
Page 12: ...1 General Information ...
Page 18: ...2 Probing ...
Page 29: ...2 12 ...
Page 30: ...3 The Format Menu ...
Page 42: ...4 The Trigger Menu ...
Page 79: ...4 38 ...
Page 80: ...5 The Listing Menu ...
Page 86: ...6 The Waveform Menu ...
Page 106: ...7 The Compare Menu ...
Page 122: ...8 The Chart Menu ...
Page 133: ...8 12 ...
Page 134: ...9 Markers and Stop Measurements ...
Page 149: ...9 16 ...
Page 150: ...10 Skew Adjust ...
Page 156: ...11 Error Messages ...
Page 165: ...11 10 ...
Page 166: ...12 Specifications and Characteristics ...
Page 176: ...13 Installation and Service ...
Page 189: ...13 14 ...
Page 193: ...Index 4 ...