Agilent 1200 Series Autosampler Reference Manual
167
Repairing the Autosampler
4
SUD Board
Frequency
When defective.
Tools required
Pozidrive No. 1 screwdriver.
Parts required
SUD board G1313-66503.
Preparations for this
procedure
Remove the top cover (
Remove the transport assembly (
Remove the sampling unit (
).
C A U T I O N
Take care not to damage the flex board when removing the SUD board.
1 Disconnect all connectors from the SUD board.
2 Remove the two fixing screws, and lift out the board.
Fixing screws