Agilent 1200 Series Autosampler Reference Manual
99
Troubleshooting and Test Functions
3
6
Check the tension of the belts.
For this use the
Torque2.mac
and measure the torque for each axis.
7
Exchange the gripper arm assembly (part number G1313-60010).
8
Adjust the belt tension.
•
If the measured torque value is too low, the belt needs to be tightened.
•
If the measured torque value is too high, the belt needs to be loosened.
For this, slide the motor (X or theta) on the holder bracket in the
appropriate direction and test the tension with the
torque2
macro. Repeat
this steps until the values are in the appropriate torque range.
9
Exchange the sample transport assembly (G1329-60009).
10
Exchange the main board (part number G1329-69520).
Table 13
Typical ranges
Theta (both) 30-50
X-axis (both) 50-90
Z-axis (both) 90-130
Gripper open 30-65
Gripper closed maximum 30
N O T E
If the Gripper open/closed torque is not in the range, proceed with STEP 7. If the theta or X
torque is not in the range, proceed with STEP 8 (if you think you can adjust the torque),
otherwise proceed with STEP 9.
Summary of Contents for 1200 series
Page 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Page 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Page 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Page 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Page 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Page 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Page 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Page 277: ......