Agilent 1200 Series Autosampler Reference Manual
147
Repairing the Autosampler
4
Injection-Valve Assembly
Frequency
When defective
Tools required
Pozidrive No. 1 screwdriver
1/4 inch wrench
Parts required
Injection valve 0101-0921 for G1329A
Injection valve 0101-1267 for G2260A
Preparations for this
procedure
Remove all capillaries from the injection valve (
).
Remove the top cover (
Remove the transport assembly (
Remove the sampling unit (
).
1 Disconnect the valve-cable connector from the SUD board
(labelled “Valve Assy”).
2 Remove the three fixing screws.