Agilent 1200 Series Autosampler Reference Manual
139
Repairing the Autosampler
4
Transport Assembly
1
Remove the top cover, top plate, and foam (see
2
Remove the Illumination assembly as described on
3
Lift out the transport assembly. This may require a flat head screwdriver to
separate the transport assembly from the sampling unit
Frequency
Sticking or jammed transport assembly.
Defective flex board or sensors.
Tools required
If interface board installed: Flat-head screwdriver.
Parts required
Transport assembly G1329-60009 for G1329A - G2260A
Figure 18
Removing the Transport Assembly