28
Agilent 1200 Series Autosampler Reference Manual
1
Installing the Autosampler
Installing the Thermostatted Autosampler
1
Place the Thermostat on the bench.
2
Remove the front cover and route the condensation drain tube to the waste.
Preparation
Locate bench space
Provide power connection
Unpack the Sampler and the Thermostat
Parts required
Sampler and Thermostat
Power cord, for the other cables see below and
Chemstation and/or Control Module G1323B.
WA R N I N G
Make sure that the condensation tube is always above the liquid level in the vessel.
If the tube is located in liquid the condensed water cannot flow out of the tube and
the outlet is blocked. Any further condensation will then remain in the instrument.
This may damage the instruments electronics.