Repairing the Autosampler
4
Entering the Serial Number using the ChemStation
Module serial numbers are entered by typing specific commands into the
command line at the bottom of the main user interface screen.
1
To enter a module serial number, type the following command into the
command line:
print sendmodule$(lals, "ser YYYYYYYYYY")
Where: YYYYYYYYYY is the 10-character serial number of the module in
question.
The reply line will respond with RA 0000 SER followed by the module serial
number you just entered.
2
Turn OFF the autosampler, then on again. Then, re-boot the ChemStation. If
the serial number you have just entered is different than the original
module serial number, you will be given the opportunity to edit the
configure Agilent 1200 Series access screen during the re-boot of the
ChemStation.
3
After boot-up, the serial number you have just entered can be seen under
the
Instrument menu
of the main user interface screen. The serial number of
the autosampler can also be seen by typing the following command into the
command line:
print sendmodule$(lals, "ser?")
The reply line will give the module serial number.
N O T E
The first two characters are letters, which should be capitalized.
Summary of Contents for 1200 series
Page 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Page 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Page 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Page 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Page 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Page 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Page 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Page 277: ......