126
Agilent 1200 Series Autosampler Reference Manual
4
Repairing the Autosampler
7 Place the piston guide on top of the seal.
8 Reassemble the metering head assembly. Carefully insert
the plunger into the base. The closed side of the metering
head must be on the same side as the lower one of the two
capillary drillings.
9 Install the fixing bolts. Tighten the bolts securely.
10 Install the metering head assembly in the autosampler.
Ensure the large hole in the metering head is facing
downwards.
CellFrame
CellFrame
Summary of Contents for 1200 series
Page 1: ...A Agilent 1200 Series Standard and Preparative Autosamplers Reference Manual ...
Page 12: ...12 Agilent 1200 Series Autosampler Reference Manual Contents ...
Page 42: ...42 Agilent 1200 Series Autosampler Reference Manual 1 Installing the Autosampler ...
Page 104: ...104 Agilent 1200 Series Autosampler Reference Manual 3 Troubleshooting and Test Functions ...
Page 174: ...174 Agilent 1200 Series Autosampler Reference Manual 4 Repairing the Autosampler ...
Page 198: ...198 Agilent 1200 Series Autosampler Reference Manual 5 Parts and Materials ...
Page 232: ...232 Agilent 1200 Series Autosampler Reference Manual 7 Introduction to the Autosampler ...
Page 277: ......