Agilent 1200 Series Autosampler Reference Manual
113
Repairing the Autosampler
4
5 Reconnect the sample-loop fitting to the needle fitting.
6 Use “Needle Up” to lift the needle to a position
approximately 2mm above the seat.
7 Ensure the needle is aligned with the seat.
On completion of this procedure:
•
Install the front cover.
•
Select “End” in the maintenance function “Change
Needle” (see
CellFrame