95
2900H EN 20071108
Advanced operation (Key-operator mode)
D
RYSTAR
5500/5503 R
EFERENCE
MANUAL
6
Press the Confirm key to select ‘Look-up table’.
The ‘Look-up table’ screen for the selected modality is displayed:
7
Press the Up/Down arrow keys to select the desired lookup table.
8
Press the Confirm key to store the data, or the Escape key to quit the procedure
without any changes.
•
In case Kanamori-like is selected, change the Kanamori-value.
An additional screen for the Kanamori-like value is displayed:
All Kanamori-like curves are selectable starting from 75 until 220 in steps of 1 unit.
The activated setting is marked with an *.
In case a number of settings for a modality were set via a remote computer, you will
not be able to change the image quality information for this modality manually. If
this is the case, the following screen will appear:
In case the image quality parameters need to be changed for this modality, please
use the DICOM Configuration tool, in the Specialist or Service mode (specialist
mode available only via remote network PC). You can also delete this modality
(refer to
) and create a new modality with the
same name. For this new modality new parameters can be assigned via the local
keypad.
<Modality name>
1 Kanamori
2 Linear
3 Kanamor-like (100)
<Modality name>
100
<Modality name>
Summary of Contents for Drystar 5500
Page 1: ...Drystar 5500 5503 Reference manual 2900H EN 20071108 ...
Page 6: ...6 2900H EN 20071108 DRYSTAR 5500 5503 REFERENCE MANUAL ...
Page 226: ...226 2900H EN 20071108 Troubleshooting DRYSTAR 5500 5503 REFERENCE MANUAL ...
Page 227: ...Equipment information sheet Appendix A ...
Page 237: ...Drystar media density response data Appendix B ...
Page 241: ...Quality Control Charts Appendix C ...
Page 252: ...252 2900H EN 20071108 Quality Control Charts DRYSTAR 5500 5503 REFERENCE MANUAL ...
Page 253: ...Remarks for HF emission and immunity Appendix D ...
Page 259: ...Index Chapter E ...
Page 265: ...265 2900H EN 20071108 Index DRYSTAR 5500 5503 REFERENCE MANUAL ...