147
2900H EN 20071108
Advanced operation (Key-operator mode)
D
RYSTAR
5500/5503 R
EFERENCE
MANUAL
2
To determine the reference values for geometry, measure the distances A and B
of the geometric square on the test image.
3
Record these values as reference dimensions A
ref
and
B
ref
on Chart 4
(‘Drystar 5500/5503 Geometric Consistency Control Chart’). Refer to
on page
.
These charts will be used for the annual quality test. For more information, refer to
‘Performing the Annual QC tests for general radiography applications (DT 2 B and
DT 2 C)’
on page
4
Save this film for future reference.
Make sure to measure distance A from the left edge of the left line to the
right edge of the right line and distance B from the upper edge of the
upper line to the lower edge of the lower line.
We strongly recommend using a 30 cm (12 inch) machinist scale with
0.5 mm divisions (1/64 inch).