145
2900H EN 20071108
Advanced operation (Key-operator mode)
D
RYSTAR
5500/5503 R
EFERENCE
MANUAL
9
Record the respective aim (average) values as the ‘Operating levels’ on Charts
2a and 2b (‘Drystar 5500/5503 Daily Density Control Chart’). Refer to
on page
.
The calculated ‘Operating levels’ should be as follows:
10
These charts will be used for the daily quality test. For more information, refer
to
‘Performing the daily QC test for general radiography applications (DT 2 B and
on page
Operating Level
Value (Macbeth units)
(according IEC 1223-2-4 or better)
Low density
0.4 ± 0.05
Mid density
1.2 ± 0.15
High density
2.0 ± 0.20