158
2900H EN 20071108
Advanced operation (Key-operator mode)
D
RYSTAR
5500/5503 R
EFERENCE
MANUAL
Establishing the image geometry reference values for mammography appli-
cation (DT 2 Mammo) (optional)
To establish the image geometry reference values, proceed as follows:
1
Print the QC mammography test image or use the previously printed test image.
You should obtain an image looking like this (without the dimensions A and B):
QC mammography test image
2
To determine the reference values for geometry, measure the distances A and B
of the geometric square on the test image.
Make sure to measure distance A from the left edge of the left line to the
right edge of the right line and distance B from the upper edge of the
upper line to the lower edge of the lower line.
We strongly recommend using a 30 cm (12-inch) machinist scale with 0.5
mm divisions (1/64 inch).
Low
Low
Base + Fog
Base + Fog
Max (100%)
Max (100%)
High
High
Mid
Mid
Identification
Identification
Identification
Identification
Initials: __________
Initials: __________
Initials: __________
Initials: __________
Base + Fog: ______________
Base + Fog: ______________
Base + Fog: ______________
Base + Fog: ______________
Low Density: _____________
Low Density: _____________
Low Density: _____________
Low Density: _____________
Mid Density: _____________
Mid Density: _____________
Mid Density: _____________
Mid Density: _____________
High Density: ____________
High Density: ____________
High Density: ____________
High Density: ____________
Max Density: _____________
Max Density: _____________
Max Density: _____________
Max Density: _____________
Density
Difference (High - Low): ___________
Density
Difference (High - Low): ___________
Density
Difference (High - Low): ___________
Density
Difference (High - Low): ___________
Date: __________
Date: __________
Date: __________
Date: __________
Time: __________
Time: __________
Time: __________
Time: __________
Daily density tests
Daily density tests
Daily density tests
Daily density tests
Weekly spatial resolution test
Weekly spatial resolution test
Weekly spatial resolution test
Weekly spatial resolution test
Quality Control Test Image - Mammo version
(Version 4
)
Sets of dots visible? Yes No
Sets of dots visible? Yes No
Sets of dots visible? Yes No
Sets of dots visible? Yes No
Dimension B
Dimension A
1
2
Summary of Contents for Drystar 5500
Page 1: ...Drystar 5500 5503 Reference manual 2900H EN 20071108 ...
Page 6: ...6 2900H EN 20071108 DRYSTAR 5500 5503 REFERENCE MANUAL ...
Page 226: ...226 2900H EN 20071108 Troubleshooting DRYSTAR 5500 5503 REFERENCE MANUAL ...
Page 227: ...Equipment information sheet Appendix A ...
Page 237: ...Drystar media density response data Appendix B ...
Page 241: ...Quality Control Charts Appendix C ...
Page 252: ...252 2900H EN 20071108 Quality Control Charts DRYSTAR 5500 5503 REFERENCE MANUAL ...
Page 253: ...Remarks for HF emission and immunity Appendix D ...
Page 259: ...Index Chapter E ...
Page 265: ...265 2900H EN 20071108 Index DRYSTAR 5500 5503 REFERENCE MANUAL ...