254
2900H EN 20071108
Remarks for HF-emission and immunity
D
RYSTAR
5500/5503 R
EFERENCE
MANUAL
Remarks for HF-emission and immunity
This device is intended for operation in the electromagnetic
environment given below. The user of the device should ensure
that it is used in such an environment.
This device was tested for a normal hospital environment as described above.
Nevertheless the HF-emission and immunity can be influenced by connected
data cables depending on length and the manner of installation.
Transmission Measurements
Agreement
Electromagnetic Environment
Guidelines
High frequency transmissions in
accordance with CISPR 11
Group 1
The device uses high frequency energy
exclusively for its internal functions. For this
reason, its high frequency transmission is
very low and it is improbable that
neighboring electronic equipment will be
disrupted.
High frequency transmissions in
accordance with CISPR 11
Class A
The device is intended for use in an
environment that do not include living areas
and in areas directly connected to a public
supply network that also supplies buildings
that are used for domestic purposes
Excess oscillations in accordance
with IEC 61000-3-2
Class A
Voltage fluctuations / flickering in
accordance with IEC 61000-3-3
Fulfilled
Summary of Contents for Drystar 5500
Page 1: ...Drystar 5500 5503 Reference manual 2900H EN 20071108 ...
Page 6: ...6 2900H EN 20071108 DRYSTAR 5500 5503 REFERENCE MANUAL ...
Page 226: ...226 2900H EN 20071108 Troubleshooting DRYSTAR 5500 5503 REFERENCE MANUAL ...
Page 227: ...Equipment information sheet Appendix A ...
Page 237: ...Drystar media density response data Appendix B ...
Page 241: ...Quality Control Charts Appendix C ...
Page 252: ...252 2900H EN 20071108 Quality Control Charts DRYSTAR 5500 5503 REFERENCE MANUAL ...
Page 253: ...Remarks for HF emission and immunity Appendix D ...
Page 259: ...Index Chapter E ...
Page 265: ...265 2900H EN 20071108 Index DRYSTAR 5500 5503 REFERENCE MANUAL ...