HWS.1 Handy Wire Scanner - 240HW01V00 

B2798_E 1.1, 2016-04-08 

 

17 

When the instrument is in operation, the state of charge of the battery is indicated by a special icon on the display 

 PLEASE REFER TO CHAPTER 2.12 DISPLAY 

 

When you try to turn on the instrument and the state of charge of the battery is low, the battery LED indicator lights 

up with three long flashes. 

 

To charge the battery use the power supply and USB cable provided: 

 

 

Check that the voltage indicated on the identification plate corresponds to the mains voltage available. 

 

Plug the charger into the socket and connect the USB cable to the corresponding connector USB-A 

 

Open the door under the handle of the instrument and connect the USB cable to the corresponding 

connector USB-B 

 

The charging process starts as soon as the charging cable is connected to the jack. 

 

The charging indicator LED under the display flashes to indicate that the battery is charging. 

 

After charging the LED indicator remains lit 

 

The charging time with the charger supplied (1500 mA) is about 3 hours.  

 

To be recharged, HWS.1 can also be connected to the USB socket of a PC (500 mA), in this case the duration of the full 

charge is approximately 8 hours.  

 

During charging, if the instrument is turned on, the icon of the state of charge of the battery indicates the type of 

charging in progress 

 PLEASE REFER TO CHAPTER 2.12 DISPLAY 

 

 

Attention! 

During the charging process the handle can be heated a bit, but this is completely normal.

 

 

 

Attention ! 

If battery charging is not possible, check that there is mains voltage on the socket; that there is good 

contact along the charging cable

 

 

In use, when the battery charge power is low, the device will automatically turn off after a warning message on the 

display (5 seconds). 

 

 

2.9

 

TURNING ON AND OFF 

 

To turn the instrument on, 

set the Safe/Arm in position A

 (Armed) and 

push the trigger for more than 3 seconds

: a 

long beep will accompany the lighting of the display. 

 

At the start up, the instrument performs an 

initialization/self-calibration procedure of the

 CCD (about 3 seconds) 

during which there must be NO object in the measurement field and NO dirt on the optical windows.  

We suggest to clean the optics before switching the device on  

If the initialization is not performed correctly for one of these reasons, on the display will show an error message 

INIT

 

CCD FAIL! REMOVE PIECE FROM THE FIELD. 

In this case,  remove  the piece from the  measuring field  and  press  a 

function key to restart the instrument. 

 

At startup is also made a test to find dirt on the optical windows. If dirt is detected the following message appears on 

the display 

DIRTY DETECTED POWER OFF & CLEAN

 

At this point, using the function keys, you can choose: 

 

F1 - Ignore  

 Start the gun ignoring the message of dirt 

F4 - Power Off  

 Turn off the instrument for cleaning of optical windows 

 

If you decide to ignore the message about the dirt, the gun will start normally, but the measurements may be not 
accurate. In this case the display will show the icon 

. The number on the icon indicates the type of detected dirty:  

1 = smooth dirty, 2 = localized dirty, 3 = smooth and localized dirty.

 

 

To switch off, press and hold (> 2 sec) the 

F3 key

 (

Power Off

) and then the 

F1 key

 as indicated on the display. 

 

 

Attention! 

In case of malfunction of the gun, the instrument can be reset by pushing the trigger for 45 seconds, 

ignoring any error messages that appear on the screen.

 

Summary of Contents for HWS.1

Page 1: ...eroel it HWS 1 Handy Wire Scanner USER S MANUAL ENGLISH HW 240HW01V00 SW 600V00L02 B2798_E 1 1 VALIDAZIONE ANNULLAMENTO numero DTE B2798_E 1 0 foglio 1 di tot 36 Rif 240HW01V00 Redatto PS Ai sensi di...

Page 2: ......

Page 3: ...2 5 FIXTURE FOR THE ON LINE MEASUREMENT OPTIONAL 13 2 6 PULLEY SET FOR WIRE GUIDANCE WITH ADJUSTING WHEEL OPTIONAL DEVICE 14 2 7 FRONT PULLEY FOR WIRE GUIDANCE OPTIONAL DEVICE 16 2 8 CHARGING THE BATT...

Page 4: ...AEROEL s r l PRADAMANO UD ITALY 4 B2798_E 1 1 2016 04 08...

Page 5: ...component interface or parts in general sold but not manufactured by AEROEL e g Personal Computers Interfaces Printers and other peripherals etc will be covered by 12 months warranty starting from th...

Page 6: ...is advisable to use the following general precautions Work Area Safety Keep your work area clean and well lit Cluttered or dark areas invite accidents Do not operate equipment in explosive atmosphere...

Page 7: ...e accessories and protections described in the following chapters of this manual in any case avoid contact even accidentally with the product or moving parts Avoid measuring moving products if the typ...

Page 8: ...bjects The following uses are prohibited use of the HWS 1 portable micrometer in an environment other than air use of the HWS 1 portable micrometer in areas classified according to the directive 99 92...

Page 9: ...ce with the laws in force in matters of waste disposal without any additional cost to the Customer except for transport costs from the Customer s site to Aeroel s premises which will be at the Custome...

Page 10: ...d by 2 CCD sensors and allows the measurement of the diameter of the product The software YesWeGun for PC Windows is available in the CD ROM supplied with HWS 1 The program allows the management of HW...

Page 11: ...in parts Display Function keys Functio UBS Connector Upper optics Lower optics Handle Product being measured Safe Arm Trigger USB Connector Info and CE marking Fixing holes Display Function keys LED b...

Page 12: ...protection is also guaranteed by the holes in which the reference spheres are inserted see figure at right To unlock the protection just pull it out and remove it from the magnet blocks This protectio...

Page 13: ...o be fixed on the machine the holder contains a TAG that identifies the machine and which is read by the instrument at the time of measurement The support allows to easily and quickly place the scanne...

Page 14: ...wire guiding system is composed by 1 rail type ITEM 6 30x30 ITEM 0 0 419 01 Assembly of 2 ceramic pulleys and 1 pre setting wheel with 6 pre setting positions from 0 5 to 5 6 mm wire diameter The whe...

Page 15: ...sembly using the nuts sliding inside the rail groove Position the pulley assembly in order to apply a slight tension to the wire the pulley should rotate when the wire moves Now lock the assembly to t...

Page 16: ...he break of the instrument the dragging of the gun the expulsion and thrown of objects The front pulley are simply fixed to the instrument through the magnets that are located on the external faces of...

Page 17: ...along the charging cable In use when the battery charge power is low the device will automatically turn off after a warning message on the display 5 seconds 2 9 TURNING ON AND OFF To turn the instrum...

Page 18: ...3 and the Enter key F1 2 11 MAINTENANCE For the right maintenance you have to do when needed the cleaning of of the deflecting prisms This operation has to be done by using a clean cloth eventually da...

Page 19: ...e status 500 mA battery charger USB standard connected battery charge status PLEASE REFER TO CHAPTER 2 8 CHARGING THE BATTERY Dirty Optics Icon The icon is active when the instrument is used despite a...

Page 20: ...NING Labels of the function keys They indicate the menu or action that is obtained by pressing the respective function key from F1 to F4 In this case pressing F1 you enter the Setup menu of the instru...

Page 21: ...isplay shows the icon Glass Logic Option to be activated for the measurement of transparent product PLEASE REFER TO CHAPTER 3 4 GLASS LOGIC Selecting the item with the Enter key F1 you access the menu...

Page 22: ...y month year YYYY MM DD year month day Select the desired value using the arrow keys F2 and F3 and the Enter key F1 Time Change Selecting the item with the Enter key F1 you access the window for setti...

Page 23: ...ing options English Italiano Select the desired value using the arrow keys F2 and F3 and the Enter key F1 Calibration It is used to correct the calibration of the instrument setting a fixed offset val...

Page 24: ...g the averaging mode PLEASE REFER TO CHAPTER 3 3 AVERAGING MODE To press the trigger the Safe Arm must be in position A Attention The trigger is magnetically operated avoid operating in the presence o...

Page 25: ...fferent colors This horizontal bar moves vertically on the display in relation to the horizontal position of the product if the product moves outside the measurement field the bar moves upwards if the...

Page 26: ...e vertical centering status of the product by splitting the box in which the measurement value appears into two segments The display does not show whether this misalignment is above or below as can be...

Page 27: ...ing the instrument horizontally clockwise or anticlockwise The horizontal inclination between the instrument and the product is greater than 7 5 the arrows on the display request to rotate the instrum...

Page 28: ...ke a measurement comparable with laser measuring instruments installed in the production line performing average measurements not at a single point on the product 3 4 GLASS LOGIC By enabling the logic...

Page 29: ...ication tag optional 3 Press the trigger while the product enters the measurement field the instrument will emit a short beep 4 Keep the trigger pressed and position the product correctly in the measu...

Page 30: ...TO CHAPTER 2 5 FIXTURE FOR THE ON LINE MEASUREMENT OPTIONAL Attention to measure moving products PLEASE REFER TO CHAPTER 1 6 SPECIFIC SAFETY PROVISIONS Correctly mounting the support on the line allo...

Page 31: ...diameter of a master in the measurement field Selecting the entry opens a window where the diameter of the master in the measurement field is displayed sampling time 1 second with the resolution of 0...

Page 32: ...F2 and F3 and the Enter key F1 To exit restore factory calibration resetting the offset press Exit F4 or select the entry Cancel and press Enter F1 When the factory calibration is restored the value...

Page 33: ...HWS 1 Handy Wire Scanner 240HW01V00 B2798_E 1 1 2016 04 08 33 4 APPENDIX 4 1 FIXTURE DIMENSIONS...

Page 34: ...he equipment To do this follow these instructions Position the gun as shown and remove the adhesive metal plate HWS 1 Remove the 4 screws shown in the figure Loosen the screw shown in the figure and s...

Page 35: ...HWS 1 Handy Wire Scanner 240HW01V00 B2798_E 1 1 2016 04 08 35 For a possible repositioning of the shell you must perform the previous steps in reverse order Remove the battery out of its housing...

Page 36: ...AEROEL s r l PRADAMANO UD ITALY 36 B2798_E 1 1 2016 04 08...

Reviews: