70
thermometer will use the RX/RS ratio measurement mode, and REF2 needs to access the standard
resistor;
Thermocouple break test: Check if the thermocouple is disconnected or not
Cold junction type: fixed and custom
Fixed value of Cold junction : the default is 0 °C, the settings can be manually input in (-20~100) °C
Wiring length(0-50)m
2.4.2 Display Settings
Display setting is the same as the
shown sheet of multi-channel thermometer
2.4.3 Temperature Measurement Process
Users can directly access two standard platinum resistance(SPRT), thermal resistance or thermistors for
temperature measurement
Users can also access standard thermocouples and industrial thermocouple for temperature measurement.
There are two ways for cold junctions compensation of thermocouples: