49
DC Current Accuracy
Test Range
Scanning Speed
Resolution
24 hours
(23 ±1) °C
90 days
(23 ±5) °C
1 year
(23 ±5) °C
Burden
Voltage
(-100-100) µA
Slow Speed
0.01 nA
15 ppm + 3 ppm
50 ppm + 6 ppm
<1 mV
8 ppm + 0.1 nA
Medium Speed
0.01 nA
15 ppm + 7 ppm
50 ppm + 10 ppm
Fast Speed
0.1 nA
15 ppm + 23 ppm
50 ppm +26 ppm
(-1-1) mA
Slow Speed
0.1 nA
15 ppm + 0.6 ppm
50 ppm +1 ppm
<1 mV
8 ppm + 0.5 nA
Medium Speed
0.1 nA
15 ppm + 1.6 ppm
50 ppm + 2 ppm
Fast Speed
1 nA
15.6 ppm + 3.6ppm
50.6 ppm + 4 ppm
(-10-10) mA
Slow Speed
1 nA
30 ppm + 3 ppm
75 ppm + 6 ppm
<1 mV
8 ppm + 10 nA
Medium Speed
1 nA
30 ppm + 7 ppm
75 ppm + 10 ppm
Fast Speed
10 nA
30 ppm + 23 ppm
75 ppm + 26 ppm
(-100-100) mA
v
10 nA
40 ppm + 0.6 ppm
75 ppm + 1 ppm
<1 mV
8 ppm + 50 nA
Medium Speed
10 nA
40 ppm + 1.6 ppm
75 ppm + 2 ppm
Fast Speed
100 nA
40.6 ppm + 3.6 ppm
75.6 ppm + 4 ppm
* Accuracy index: ± (ppm r ppm range), choose the largest
* Temperature Coefficient index:Exceed the range of range of (18-28) °C, increase (ppm xxnA)/°°C.
* Input Protection 0.3A/600V Resettable PTC.
Accuracy Rating of DC Resistance
Summary of Contents for ADT286
Page 1: ......
Page 20: ...19 1 3 Base Structure Picture 1 ADT286 Mainframe...
Page 22: ...21 Picture 2 Connection of inserting and unplugging of embedded scanners...
Page 23: ...22 1 3 1 ADT286 TS Multi channel temperature signal scanner Picture 3 Temperature Device...
Page 25: ...24 1 3 2 ADT286 PS Multi channel process signal scanner Picture 4 Process Device...
Page 129: ......