34 

 

1.4.4 ADT280-25R/100R Standard Resistor 

 This resistor is mainly used for the RX/RS resistance ratio measurement mode of REF2 channel on two 

front panel of ADT286 

   To provide resistance values 

of 25Ω and 100Ω 

 Simply insert into the front panel   

 

1.5 Service Environment 

 Working environment temperature range: (0~50) °C (guarantee accuracy:18 °C ~ 38 °C) 

 Storage environment temperature range: (-20~70) °C 

 Ambient humidity: 0~90% RH (0°C~50°C), non-condensing 

 

Summary of Contents for ADT286

Page 1: ......

Page 2: ...1 Additel 286 Multifunction Reference Thermometer Readout User s Manual Version 201911V01 Additel Corporation...

Page 3: ...n Reference Thermometer Readout To ensure correct operation and safety please follow the instructions in this manual Additel Corporation reserves the right to change the contents and other information...

Page 4: ...ent wiring device 26 1 3 4 ADT280 25R 100R standard resistor 28 1 3 5 Connection of ADT286 TS PS 29 1 4 Features 30 1 4 1 ADT286 TS Multi channel temperature signal scanner 31 1 4 2 ADT286 PS Multi ch...

Page 5: ...hermometer 57 2 2 1 Channel Switch 58 2 2 2 Measuring Project 59 2 2 3 Channel Settings 60 2 2 4 Display settings 65 2 3 DAQ 66 2 3 1 Scanner 66 2 3 2 Align Channel 67 2 3 3 Channel Settings 67 2 3 4...

Page 6: ...Modification 82 3 1 2 Sensor Library 82 3 1 3 Standard Resistor Library 82 3 1 4 Data Management 82 3 1 5 Application Configuration Files 82 3 1 6 System Service 82 3 2 Sensor Library 82 3 2 1 Industr...

Page 7: ...ting 90 3 4 1 Ethernet 91 3 4 2 Wireless Communication 92 3 5 Personalization 94 3 5 1 Date and Time 95 3 5 2 Language 95 3 5 3 Sound 96 3 5 4 Screen Contrast 96 3 6 System Service 96 3 6 1 System Cal...

Page 8: ...on 104 4 1 Probe Calibration 104 4 1 1 Data Managemnet 105 4 1 2 Display Settings 105 4 1 3 Operation Process 106 4 2 Environmental Temperature Field Test 111 4 2 1 Operation Procedure 112 4 2 2 Data...

Page 9: ...116 4 4 3 Dry Well Axial Uniformity Test 117 4 4 4 Bath Uniformity Test 118 4 4 5 Furnace Radial UniformityTest 119 4 4 6 Furnace Axial Uniformity Test 120 4 5 Snapshot 121 4 5 1 How to Set 122 4 5 2...

Page 10: ...canner 62 Sheet 12 channel settings of process scanner 63 Sheet 13 Connection Settings 65 Sheet 14 General function of DC multimeter settings menu 76 Sheet 15 Standard Thermocouple 83 Sheet 16 Standar...

Page 11: ...10 Sheet 27 Acloud Service 104 Sheet 28 Screen shot 122 Sheet 29 Thermo Technical Calculator of Thermocouple 125 Sheet 30 Thermotechnical Calculator of Thermal Resistance 126...

Page 12: ...ury Please do as follows Please read the instructions manual before using the product The thermometer must be operated by trained personnel to prevent injury or damage to the operator or thermometer B...

Page 13: ...gnated test leads Please disconnect the test cable before switching the measurement gear Do not disassemble the product without instruction from Additel Attention To prevent injury please obey the ins...

Page 14: ...meter For cleaning of the thermometer please consult Additel It is strictly forbidden to use fuses that do not meet the requirements of the corresponding thermometer model If the thermometer is abnorm...

Page 15: ...ement temperature and humidity filed testing DMM and other functions which provides users with the high temperature measurement accuracy flexible configurations easy probe wiring channel expansion and...

Page 16: ...eter Single or multi channel measurements DAQ Data acquisition DMM DC digital multimeter Smart Connection Remote wiring connections Probe Calibration Calibration and characterizations of various probe...

Page 17: ...pture Thermal Calculator Signal calculation for various sensor types Intelligent communication of temperature Sources Control and interface with various temperature sources Sensor Library Stores speci...

Page 18: ...annels of 2 wire or 10 channels of 3 4 wire which can be used to switch signals of thermal resistance thermocouple thermistor DC voltage and DC resistance 3 ADT286 PS Multichannel Process Signal Scann...

Page 19: ...DSUB 3M DSUB 10M 3 or 10 meter DSUB connection cable This cable is used scannerfor remote or cascading connection to the temperature and or process scanner a 9 DG01 Scanner cover scannerImproves meas...

Page 20: ...19 1 3 Base Structure Picture 1 ADT286 Mainframe...

Page 21: ...1inch TFT capacitive screen The main screen of thermometer 6 Power switch Power unit on and off 7 Power specification Specification of external work power 8 AC port AC socket port 9 Internet access RJ...

Page 22: ...21 Picture 2 Connection of inserting and unplugging of embedded scanners...

Page 23: ...22 1 3 1 ADT286 TS Multi channel temperature signal scanner Picture 3 Temperature Device...

Page 24: ...hannel has it s own cold junction compensation sensor 3 Grounding receptacle For ground connection 4 Amphenol connector The patented press quick Amphenol connector supports bare wire shovel MiNiTC ban...

Page 25: ...24 1 3 2 ADT286 PS Multi channel process signal scanner Picture 4 Process Device...

Page 26: ...Grounding receptacle For DUT ground connection of sensor 4 Cascade address and enable status indicator lamp To indicate the scanner is receiving power and if it is cascaded to another scanner 5 The e...

Page 27: ...26 1 3 3 ADT286 Dock Remote intelligent wiring device Picture 5 Remote intelligent wiring device...

Page 28: ...in device will be transferred to this terminal 4 Amphenol connector The signals of one channel wiring terminal in device will be transferred to this terminal 5 Display screen To indicate the channel...

Page 29: ...28 1 3 4 ADT280 25R 100R Standard Resistor Picture 6 This Standard Resistor is inserted into REF2 channel Picture 6 Standard Resistor...

Page 30: ...5 Standard Resistor No Name Explanation 1 Resistance plug Wiring plug 2 Resistance label With QR code resistance Value types and other information 1 3 5 Connection of ADT286 TS PS Picture 7 wiring co...

Page 31: ...an be connected to the SPRT the REF2 can be connected to the external standard resistor adopting the RX RS resistance ratio measurement mode Current reversal technology is used The front panel channel...

Page 32: ...o switch the signals of thermal resistance thermocouple thermistor DC voltage and DC resistance Channel number 2 wire DUT for 20 channels 3 4 wire DUT for 10 channels The scanner can be directly embed...

Page 33: ...the signal of the 9 channel transmitter channel The transmitter is powered by 24V The supported transmitters are temperature transmitter humidity transmitter pressure transmitter etc The transmitter...

Page 34: ...itoring Before the verification test starts the remote ADT286 main unit automatically sends the detected information to the display of the device The user finds the corresponding DUT according to prom...

Page 35: ...e of REF2 channel on two front panel of ADT286 To provide resistance values of 25 and 100 Simply insert into the front panel 1 5 Service Environment Working environment temperature range 0 50 C guaran...

Page 36: ...ng pulley 100V 120V 220V 240V 36Wmax 47Hz 440Hz Screen Specification 10 1 inch TFT color screen capacitive touch screen Communication Interface USB A USB B LAN WIFI Bluetooth Language Chinese English...

Page 37: ...e Temperature signal scanner 3 4 wire for 10 channels or 2 wire for 20 channels Process signal scanner 1 standard reference channel 9 transmitters or switch channels Electrical Specifications1 hour Wa...

Page 38: ...or 0 4m1 1 mA 12 V 3 ppm 0 02 m Medium Speed 0 01 m 3 ppm or 0 7 m 12 ppm or 0 8 m 15ppm or 0 8m1 Fast Speed 0 1 m 3 6 ppm or 1 9 m 12 6 ppm or 2 m 15 6ppm or 2m5 0 4000 Slow Speed 0 1 m 3 ppm or 4 m...

Page 39: ...600 C 0 0029 0 0146 0 0146 0 0030 Medium Speed 200 C 0 0013 0 0016 0 0016 0 0002 0 C 0 0014 0 0031 0 0038 0 0008 300 C 0 0020 0 0089 0 0089 0 0018 600 C 0 0029 0 0146 0 0146 0 0030 Fast 200 C 0 0039 0...

Page 40: ...nel Channel REF1 RX REF2 RS Range Reference Resistance Ratio Rx Rs 1 Year 23 5 C ppm of Reading 100 25 2 00 4 00 1 5 1 10 2 00 0 85 0 90 1 10 0 6 0 50 0 90 1 5 0 25 0 50 2 5 400 100 2 00 4 00 2 1 10 2...

Page 41: ...2 57 1 5 1 11 660 323 3 37 1 5 1 58 PT100 100 189 3442 0 22 1 2 0 07 38 8344 0 84 0 95 0 20 0 01 1 0 26 0 07 231 928 1 89 0 81 0 42 419 527 2 57 2 1 47 660 323 3 37 2 2 11 The PT25 indicator is based...

Page 42: ...m 10 ppm 80 m 30 ppm 80 m 40 ppm 80 m Fast Speed 100 m 10 6 ppm 200 m 30 6 ppm 200 m 40 6 ppm 200 m 100 1000 k Slow Speed 0 1 50 ppm 80 ppm 100 ppm 1 A 5 ppm 1 Medium Speed 0 1 50 ppm 1 80 ppm 1 100 p...

Page 43: ...Medium Speed 40 C 0 0007 0 0011 0 0014 0 C 0 0002 0 0006 0 0008 50 C 0 0008 0 0010 0 0011 100 C 0 0054 0 0064 0 0064 150 C 0 0239 0 0282 0 0282 Fast Speed 40 C 0 0007 0 0011 0 0014 0 C 0 0002 0 0006...

Page 44: ...1 V Medium Speed 0 01 V 5 ppm 6 ppm 10 ppm 8 ppm 14 ppm 8 ppm Fast Speed 0 1 V 5 ppm 22 ppm 10 ppm 24 ppm 14 ppm 24 ppm Accuracy Index ppm of reading ppm of FS Temperature coefficient index Exceed the...

Page 45: ...31 0 014 0 009 0 036 0 020 0 015 0 039 0 022 0 017 1000 0 034 0 017 0 012 0 042 0 025 0 019 0 046 0 029 0 024 J 200 0 102 0 043 0 025 0 113 0 054 0 036 0 115 0 055 0 037 100 0 054 0 022 0 013 0 060 0...

Page 46: ...0 022 0 016 400 0 037 0 016 0 010 0 042 0 021 0 015 0 044 0 023 0 016 R 40 0 543 0 222 0 124 0 593 0 272 0 173 0 593 0 272 0 173 0 0 416 0 170 0 095 0 454 0 208 0 132 0 454 0 208 0 132 155 0 266 0 10...

Page 47: ...71 0 038 0 028 1000 0 062 0 028 0 018 0 072 0 038 0 028 0 075 0 042 0 031 1200 0 065 0 030 0 019 0 076 0 041 0 031 0 081 0 046 0 035 1300 0 068 0 032 0 020 0 080 0 044 0 033 0 085 0 049 0 038 L 200 0...

Page 48: ...ccuracy Test Range Scanning Speed Resolution 24 hours 23 1 C 90 days 23 5 C 1 year 23 5 C Input Resistance Temperature Coefficient 100 100 mV Slow Speed 0 01 V 5 ppm 2 ppm 10 ppm 4 ppm 14 ppm 4 ppm 10...

Page 49: ...eed 10 V 8 ppm 1 ppm 32 ppm 1 ppm 38 ppm 1 ppm 10 M 5 ppm 5 V Medium Speed 10 V 8 ppm 2 ppm 32 ppm 2 ppm 38 ppm 2 ppm Fast Speed 100 V 8 6 ppm 7 ppm 32 6 ppm 7 ppm 38 6 ppm 7 ppm Accuracy Index ppm of...

Page 50: ...2 ppm Fast Speed 1 nA 15 6 ppm 3 6ppm 50 6 ppm 4 ppm 10 10 mA Slow Speed 1 nA 30 ppm 3 ppm 75 ppm 6 ppm 1 mV 8 ppm 10 nA Medium Speed 1 nA 30 ppm 7 ppm 75 ppm 10 ppm Fast Speed 10 nA 30 ppm 23 ppm 75...

Page 51: ...3 ppm 15 ppm 0 3 ppm 1mA 3ppm 0 02m3 Medium Speed 0 1 m 3 ppm 1 2 ppm 12 ppm 1 3 ppm 15 ppm 1 3 ppm Fast Speed 1 m 3 6 ppm 3 2 ppm 12 6 ppm 3 3 ppm 15 6 ppm 3 3 ppm 0 10 k Slow Speed 1 m 3 ppm 0 3 pp...

Page 52: ...m 0 100 M Slow Speed 10 150 ppm 1 ppm 400 ppm 4 ppm 500 ppm 4 ppm 0 1 A 50 ppm 50 Medium Speed 10 150 ppm 6 ppm 400 ppm 9 ppm 500 ppm 9 ppm Fast Speed 100 150 ppm 11 ppm 400 ppm 14 ppm 500 ppm 14 ppm...

Page 53: ...Sheet 6 Fuse Specification Power Supply Requirement 90V 110V or 108V 132V 198V 242V or 216V 264V 47 440 Hz Power Dissipation 40VA peak 30Watt average Fuse specification 110V 0 315A F 250V 230V 0 315A...

Page 54: ...e is installed in it 3 Protective cover Fuse cover Display and Function Operation 2 1 Main Interface Display The default main interface consists of three parts status bar tool bar and main menu of sin...

Page 55: ...e green icon indicates that the connection is OK but without calibration and the blue icon indicates normal status 2 1 2 Menu Function Zone Lock Screen After clicking the lock screen button except for...

Page 56: ...ement can be performed on multiple channels according to the scan setup DC Multimeter The front panel REF1 channel is used as a multimeter The signals that can be measured are DC voltage DC current an...

Page 57: ...device System Settings System settings include data security communication settings personalization system services system information sensor libraries standard resistor library and the setup and mai...

Page 58: ...utomatically The default main interface displays the current scanning channel measuring type measuring range measured value scanning speed and smoothing and other information When the user selects the...

Page 59: ...58 Picture10 Multi channel thermometer graphical display 2 2 1 Channel Switch Manually Switchover Click the left and right direction buttons to switch the measuring channel or...

Page 60: ...ly scanning by time our count After settings are complete click to begin automatically scanning and the automatically scanning function will begin to scan all online channels based on user s settings...

Page 61: ...the current channel Function Measurement function Contains voltage current resistance TC Standard TC RDT SPRT Custom RTD Thermistor and other measurement items After selecting the measurement item th...

Page 62: ...JC type is set to custom and REF 2 is the designated channel REF 2 has a SPRT selected and the SPRT is in the ice bath When measurements are made for the Type S TC it will use the SPRT measurement as...

Page 63: ...4 wire configuarations Open Circuit Detect ON OFF Choosing the relative measuring items of thermocouple to indicate that connection is normal or not CJC Type Internal External Custom Internal will us...

Page 64: ...annel 1 is typically used to measure a reference standardand other channels are as DUT Sheet 10 channel settings of process scanner settings operation explanation Enabled ON OFF Whether to display the...

Page 65: ...nput 20 100 Choosing thermocouple to display Cold junction fixed value the default cold junction value is 0al will use the temperaturen 20 100 Open Circuit Detect ON OFF Only for 01 channel choosing r...

Page 66: ...pe set to Custom After the completion of function settings click the button in the right tool bar to save settings 2 2 4 Display settings Display settings include temperature unit sampling rate displa...

Page 67: ...ween the numeric display mode and the graph display mode The icon provides the function of clearing historical data 2 3 DAQ The DAQ feature allows for data recording of all channels according the conf...

Page 68: ...cord name operator and notes 2 3 2 Channel alignment In the channel alignment function the user can select one channel as the reference channel and then align the other channels to the reference chann...

Page 69: ...n is started there are various view for the display Data Chart Monitor and Info Each view can be selected on the right vertical tool bar The Data view shows the channel sensor range reading and date a...

Page 70: ...this application the user can access two standard sensors to achieve high precision temperature measurement 2 4 1 Channel Settings Enabled ON OFFscanner Label channel label name Function A sensor can...

Page 71: ...nually input in 20 100 C Wiring length 0 50 m 2 4 2 Display Settings Display setting is the same as the 2 2 4 shown sheet of multi channel thermometer 2 4 3 Temperature Measurement Process Users can d...

Page 72: ...selecting channel settings in the multi channel thermometer menu then select and set REF2 as the CJC For ratio measurement mode the user can connect the REF1 channel to the SPRT the REF2 to the exter...

Page 73: ...onfiguration information of each channel on the ADT286 main unit After editing the information of each channel is automatically sent to ADT286 Dock remote intelligent junction box and the user follows...

Page 74: ...1 Local Wiring How to operate 1 Click the scanning box shown in picture 12 or the lower right corner of the right menu bar to enter the local wiring status 2 Click the setting in the toolbar on the r...

Page 75: ...ing and now thermometer main unit will send the configuration information of remote wiring box to external intelligent wiring pedestal 4 The user finds the corresponding channel and the completed wiri...

Page 76: ...esistance Through clicking to open the measurement settings menu Click the button of measuring setting menu to close the menu display The display change icon provides switching between the numerica an...

Page 77: ...ng speed is optional in slow medium fast Display digits Display resolution offset The offset is optional When selected it provides a math null and shows the offset displayed on the main display Press...

Page 78: ...T User selectable filtering Supports fast medium slow sampling speeds Supports 6 7 8 9 display resolution 2 6 2 DCI Current Measurement Range Auto 100uA 1mA 10mA 100mA OFFSET User selectable filtering...

Page 79: ...ctable filtering Support fast medium slow sampling speeds Support 6 7 8 9 display resolution 2 6 4 Four wire resistance measurement Range Auto 100 1k 10k 100k 1M 10M 100M OFFSET User selectable filter...

Page 80: ...user can manage the connected temperature sources or connect to others The icon indicates that a temperature source can be added Click the icon to enter the temperature control source list interface...

Page 81: ...return to the connected temperature source interface after the connection is successful Edit Click on the existing temperature source device in the list to edit its communication configuration and est...

Page 82: ...tion tool can use the network or serial communication to establish a communication connection with the temperature source device to help the user to test the temperature source command System Settings...

Page 83: ...elds 3 2 Sensor Library The sensor library offers eight sensor types TC standard TC RTD SPRT custom RTD thermistor humidity transducer pressure and temperature transducers 3 2 1 Industrial Thermocoupl...

Page 84: ...ure range unit switching by clicking temperature unit Base TC sensor Letter S couple and B couple are optional Parameter type parameter type of standard thermocouple Click parameter type to make choic...

Page 85: ...mber Number letter and Chinese up to 14 digits Numeric indentification Temperature range Based on temperature unit Temperature range Negative temperature calculation coefficient Standard platinum resi...

Page 86: ...85 3 2 5 Custom RTD The user can add delete and modify the custom RTD The custom RTD includes CVD equation and RTD The RTD is the same as 3 2 3 Industrial RTD The CVD equation is as follows...

Page 87: ...ge Based on temperature unit CVD equation temperature range unit switching by clicking temperature unit R0 Digits Thermal resistance R0 value Calibration date 2000 1 1 2099 12 31 Calibration date Date...

Page 88: ...range Upper and lower limit of temperature is 273 15 9999 the default is 80 150 The default is C switchable NTC Calculation formula TC Calculation formulahablepe Different calculation parameters for d...

Page 89: ...and Chinese up to 14 digits Serial Number Transfer function Linear Square Root The default is linear switchable Input signal range 0 100 Manually input Input decimals 0 6 Manually input Output signal...

Page 90: ...2 9 Temperature Transducer The user can add delete and revise temperature transducer operation the same as sheet 19 of humidity sensor except units are in relation to temperature 3 3 Standard Resistor...

Page 91: ...Number Number letter and Chinese up to 14 digits Serial Number R digit Certificate value of standard resistor Calibration date 2000 1 1 2099 12 31 Calibration date Date of next calibration 2000 1 1 20...

Page 92: ...ss methods Items Effective value Explanation Address access DHCP Manually Choose the way to access the device address When the DHCP mode is selected the contents of the following sheet are automatical...

Page 93: ...ubnet mask 0 0 0 0 255 255 255 255 Set Subnet mask of thermometer Gateway 0 0 0 0 255 255 255 255 Set source gateway of thermometer The port number and physical address is a factory setting which can...

Page 94: ...DHCP manually Choose the ways to access the device address The port number and physical address are factory settings and cannot be changed 1 For advanced option when DHCP mode is selected the content...

Page 95: ...et mask of device Gateway 0 0 0 0 255 255 255 255 Set gateway of device Click right corner of screen to confirm the settings The wireless communication settings take effect directly without confirmati...

Page 96: ...1 Date settings Date format Year month day month day year day month year Date format settings Separator Date separator settings 3 5 2 Language The device provides a multi language interface that allow...

Page 97: ...ontrast Click the progress bar to slide left or right to adjust the screen brightness 3 6 System Service System services include system calibration maintenance factory reset and system upgrades 3 6 1...

Page 98: ...the range item which needs to be calibrated to enter calibration interface Click the button last time calibration record to check the latest calibration data Calibration process 1 Click the calibrati...

Page 99: ...the range which needs to be calibrated to enter calibration interface Click the button last time calibration record to check the latest calibration data Calibration process 1 Click the calibration sta...

Page 100: ...alibration data of resistance and restore the factory calibration Calibration Process 1 Click the calibration start button to start the calibration process after finishing the data acquisition please...

Page 101: ...ave the cold junction calibration data 3 6 2 Maintenance To enable this function you need to enter a password The default factory password is 123456 Maintenance record Users can add maintenance person...

Page 102: ...d in the calibration menu After resetting the factory settings and restarting the unit the user needs to set the time For details please refer to chapter 3 5 1 Date Time 3 6 4 System Upgrade The devic...

Page 103: ...completion information 3 7 System Information System information includes basic information version information and operation information of the thermometer 3 7 1 Basic Information Include serial NO...

Page 104: ...rowse Users can export file data through USB flash drive or PC software in CSV format Users can delete file data in batches Provide file search function to find files according to time file name opera...

Page 105: ...show the staring icon of Acloud service Application Click the application menu to enter the application function which includes probe calibration environmental temperature field test switch test const...

Page 106: ...s for Data Management and Display Settings 4 1 1 Data Management Seeing section 3 8 Data Management 4 1 2 Display Settings Temperature unit K C and F Sampling speed fast 10 times s medium speed 1 time...

Page 107: ...r to the 01 to 10 channel of the temperature scanner 4 Click to define the calibration range set points and add a temperature source 5 Click to enter the test configuration interface The thermal resis...

Page 108: ...e g 1 2 3 4 5 6 7 8 9 10 10 9 8 7 6 5 4 3 2 1 When round trip is enabled each duplicate channel measurement will be averaged This is useful when using a heatsource that is not as stable as desired Di...

Page 109: ...on for thermal resistance Click to save the data to the USB flash drive click to save the data locally and fill in the file name operator and notes when saving 2 Standard Platinum Resistance Calibrati...

Page 110: ...the users can set the number of samples filtering and sampling speed 5 Click to enter the test plan information interface set the test plan name creator and add notes 6 Click to save all configuratio...

Page 111: ...on Method Refer to the platinum RTD calibration process but note that the SPRT comparison method can select to enable the self heating effect test and the zero power tests 3 Thermocouple Calibration T...

Page 112: ...Transducer calibration The transducer calibration process can refer to the platinum RTD calibration process no coefficient calculation no standard resistance function The standard sensor can select t...

Page 113: ...ocess scanner channels Only the location and channel configured will be included in the test After the configurations of all required test points are completed the solution name and the creator should...

Page 114: ...tings 1 Function box select the function box 2 Function select standard sensor 3 Connections select reference sensorand wire configurations 4 Cold junction type Internal fixed select thermocouple as s...

Page 115: ...the switch test page for switch test 1 The CH1 01 channel of the process scanner is used to measure the reference sensor and other channels are for the DUTs 2 The switch test interface loads all the c...

Page 116: ...radial uniformity tests 4 4 1 Stability Test Test the stability of the dry well temperature over time 1 Test Settings Test duration set test time in minutes Select channel select the channel of the re...

Page 117: ...ires in the chart 4 4 2 Dry Well radial uniformity test This test measures the maximum temperature difference between different holes of the dry well inserts 1 Test settings Select channel select corr...

Page 118: ...s according to the formula Click to view the temperature difference value click to save the data 4 4 3 Axial uniformity temperature test Test the uniformity of the axial temperature distribution in th...

Page 119: ...it is stable click to take the reading then move the sensor position for the next reading After all readings are made click to save the test record 4 4 4 Thermo bath uniformity test Test the differenc...

Page 120: ...hese steps 4 times Scroll down the data field to section B position sensors correctly and take 4 meausrements and move to the next sections According to the above process when the data acquisitions of...

Page 121: ...lues two times for all position points then click to save the results 4 4 6 Furnace axial uniformity test 1 Test Settings See the test settings of dry well axial uniformity test 2 Data Management See...

Page 122: ...est At the temperature point when the furnace temperature stability meets the specified requirements click to read and when all points reading please click for save 4 5 Screen Capture The screen captu...

Page 123: ...k Select the storage position for screenshot files Storage quantity only can selecting a local storage path Read only Number of screenshots which can be saved in a local file File prefix Number letter...

Page 124: ...e screen capture list page to enter the thumbnail page of a single screen capture file You can slide all the screen captures to the left or right or click the previous button and the next button in th...

Page 125: ...rmal technical calculator provides thermocouple and thermal resistance calculation functions and also provides SPRT coefficient value calculation which is convenient for users to perform unit numerica...

Page 126: ...es of TC Temperature celsius value unit e Kelvin Based on the types of couplet Temperature Kelvin value unit K Fahrenheit degree Based on the types of couplet Temperature Fahrenheit value unit F Diffe...

Page 127: ...lectrical signal output of thermal resistance unit Celsius Based on sensor type Temperature celsius value unit t Kelvin Based on sensor type Temperature Kelvin value unit K Fahrenheit degree Based on...

Page 128: ...a5 b5 38 8433 29 7646 W Ga W Hg a4 b4 189 3442 0 01 W Hg W Ar 4 6 2 How to Use The user only needs to click the known items and input the value the thermometer will automatically calculate the value...

Page 129: ......