3 After the calibration of all calibration points is completed, click
to save the calibration data.
3.6.1.3
Resistance Calibration
Seven ranges: (0~100)Ω
、
(0~1)kr
、
(0~10)ka
、
(0~100)kn
、
(0~1)M)
、
(0~10)Mk
、
(0~100)Mn
(Including the calibration of resistance, thermal resistance and thermistor)
Click the button
to clear calibration data of resistance and restore the factory calibration
Calibration Process
1 Click the calibration start button
to start the calibration process, after finishing the data acquisition,
please click the button
to enter the next point calibration point;
2 After finish the calibration of all calibration points step by step, then click
to save calibration data.
3.6.1.4 Cold Junction Calibration
Cold junction calibration
:
Only calibration for connected temperature scanner A, B channel and process