63
the SPRT measurement as the CJC.
Cold Junction fixed Value
Manually input(-20-100)
℃
Cold junction type is fixed value display, the default cold junction
value is 0 d.em, the 1 has a Type S TC(-20-100)c
Wiring length
Manually input(0-50)m
Set the wiring length of sensor
Channel Settings of Process Scanner
The channel settings of process scanner differs from the other channel settings in that during the measurement
of process scanner, the channel 1 is typically used to measure a reference standardand other channels are as
DUT.
Sheet 10 channel settings of process scanner
settings
operation
explanation
Enabled
ON/OFF
Whether to display the measurement data
Label
Label
Customizable naming for the measurement channel
Function
Selection of measurement items
When selecting to channel 1 as a reference, you can select TC,
Standard TC, RTD, SPRT, Custom RTD. When switching to a
channel of 02~10, you can select voltage transmitter, current
transducer and temperature switch
Fixed delay/Extra delay
Manual input (0~60000)ms
The time it takes to switch from the other channel to the current