97
security
Click the button
to clear
users’ calibration data and restor the factory settings
3.6.1.1 Voltage Calibration
Four ranges: (-100~100)mV
、
(-1~1)V
、
(-10~10)V
、
(-50~50)V;
Click the button
to clean calibration data of voltage user for restoring factory calibration
Choosing the range item which needs to be calibrated to enter calibration interface
Click the
button “last time calibration record “to check the latest calibration data.
Calibration process
1 Click the calibration start button
to start the calibration process, the system will automatically collect data
until manually enter the next calibration point.
2 Click
to start calibration of the next calibration point, or click
to return to the previous calibration point
to recalibrate
Summary of Contents for ADT286